Table 5-4. CALIBRATE O2 Sub-Menu

Instruction Manual

IB-106-300NH Rev. 4.3 May 2005

World Class 3000

Table 5-4. CALIBRATE O2 Sub-Menu

SUB-MENU

SETUP SETTING

 

 

 

SELECTION

(SEE TABLE 3-5)

 

DISPLAY

DESCRIPTION

 

 

 

 

Perform

Auto Cal in Probe

Press ENTER to start Auto

MPS will start calibrating probe.

Calibration

Setup is YES

Calibration.

 

 

 

Starting Automatic Calibration

 

 

 

High Gas _____% O2

Value for high O2 calibration gas.

 

 

Time Left 0:00

Amount of time necessary to com-

 

 

 

 

plete the current testing phase in

 

 

 

 

min:sec.

 

 

Cell mV

______mV

Cell voltage of the probe.

 

 

Low Gas

_____% O2

Value for low O2 calibration gas.

 

 

Time Left 0:00

Amount of time necessary to com-

 

 

 

 

plete the current testing phase in

 

 

 

 

min:sec.

 

 

Cell mV

______mV

Cell voltage of the probe.

 

 

Resistance Check

Resistance check in progress.

 

 

Time Left 0:00

 

 

 

Cell _____mV _____C

Cell voltage and probe temperature.

 

 

Calibration Complete

Gas lines are being purged of cali-

 

 

Purging 0:00

 

 

 

 

bration gas.

 

 

Cell _____mV _____C

Cell voltage and probe temperature.

 

 

Calibration Complete

 

 

Auto Cal in Probe

Press ENTER to start Manual

Manual calibration sequence will

 

Setup is NO.

Calibration.

begin when ENTER is pressed.

 

 

Switch ON high calibration gas.

 

 

 

Press ENTER when ready.

 

 

 

High gas ______% O2

High O2 calibration gas value.

 

 

Press ENTER when O2 reading

 

 

 

is stable.

 

 

 

 

Turn OFF high calibration gas

 

 

 

and ON low calibration gas.

 

 

 

Press ENTER when ready.

 

 

 

Low gas

______% O2

Low O2 calibration gas value.

 

 

Press ENTER when O2 reading

 

 

 

is stable.

 

 

 

 

Resistance Check.

Resistance check in progress.

 

 

Turn off low calibration gas.

 

 

 

Press ENTER when ready.

 

 

 

Press ENTER when probe has

 

 

 

returned to process.

 

5-10 General User Interface (GUI) Operation