7-8 DEFINING A TIMED CALIBRATION VIA HART

Use the following procedure to specify a time interval (in hours) at which the Oxymitter 4000 will be automatically calibrated. If necessary, use the menu tree in Figure 7-3(Sheet 2 of 3) for reference.

NOTE

To select a menu item, either use the up and down arrow keys to scroll to the menu item and press the right arrow key or use the number keypad to select the menu item number. To return to a preceding menu, press the left arrow key.

a.From the DEVICE SETUP screen, select

DETAILED SETUP.

b.From the DETAILED SETUP screen, select

O2 CALIBRATION.

c.From the O2 CALIBRATION screen, select menu item 6, CAL MODE. Set the CAL MODE to AUTO.

d.Return to the O2 CALIBRATION screen and select menu item 7, CAL INTRVL.

e.At the prompt, input a time interval (in hours) at which an automatic calibration will occur and press ENTER.