Oxymitter 4000

Instruction Manual

IB-106-340 Rev. 3.0 December 2003

SECTION 9

MAINTENANCE AND SERVICE

9-1 OVERVIEW

This section identifies the calibration methods available and provides the procedures to maintain and service the Oxymitter 4000 and optional integrally mounted SPS 4000.

Install all protective equipment covers and safety ground leads after equipment repair or service. Failure to install covers and ground leads could result in serious injury or death.

9-2 CALIBRATION – OXYMITTER 4000 WITH A MEMBRANE KEYPAD

a.During a calibration, two calibration gases

with known O2 concentrations are applied to the Oxymitter 4000. Slope and constant val- ues calculated from the two calibration gases determine if the Oxymitter 4000 is correctly

measuring the net concentration of O2 in the industrial process. A calibration record sheet has been provided at the back of this section to track performance.

Before calibrating, verify that the calibration gas parameters are correct by setting the gas concentrations used when calibrating the unit (see paragraph 5-2a.5for the key- pad, or Section 6 for the LOI) and by setting the calibration gas flowmeter. The calibra- tion gas flowmeter regulates the calibration gas flow and must be set to 5 scfh. Only adjust the flowmeter to 5 scfh after placing a new diffusion element on the end of the Oxymitter 4000. Adjusting the flowmeter at any other time can pressurize the cell and bias the calibration.

In applications with a heavy dust loading,

the O2 probe diffusion element may become plugged over time, causing a slower speed of response. The best way to detect a plugged diffusion element is to note the time it takes the Oxymitter 4000 to return to the normal process reading after the last cali- bration gas is removed and the calibration

gas line is blocked off. A plugged element also can be indicated by a slightly lower reading on the flowmeter.

Change the diffusion element when the calibration gas flowmeter reads slightly lower during calibration or when the response time to the process flue gases becomes very slow. Each time the diffusion element is changed, reset the calibration gas flowmeter to 5 scfh and calibrate the Oxymitter 4000. To change the diffusion element, refer to paragraph 9-4j.

b.Three calibration methods are available: automatic, semi-automatic, and manual.

NOTE

A calibration can be aborted any time during the process. Press the CAL key (Figure 9-1)on the Oxymitter 4000 keypad three times within three seconds, or abort via the LOI, HART/AMS, or IMPS 4000. An aborted calibration retains the values of the previous good calibration.

1.Automatic Calibration. Automatic cali- brations require no operator action. However, the calibration gases must be permanently piped to the Oxymitter 4000, an SPS 4000 or IMPS 4000 must be installed to sequence the gases, and the Oxymitter 4000’s logic I/O must be set to mode 8 via HART/ AMS or the LOI so the sequencer and Oxymitter 4000 can communicate.

Depending on your system setup, an automatic calibration can be initiated by the following methods:

(a)The Oxymitter 4000’s CALIBRA- TION RECOMMENDED alarm sig- nals that a calibration is required.

(b)Enter a “time since last cal” pa- rameter (CAL INTRVL) via HART/ AMS or the LOI that will initiate an automatic calibration at a sched-

Rosemount Analytical Inc. A Division of Emerson Process Management

Maintenance and Service 9-1