Instruction Manual

IB-106-340 Rev. 3.0 December 2003

Oxymitter 4000

 

 

 

 

 

 

 

 

 

 

 

 

O2 Gas 1

_____%

(CONTINUED FROM

 

 

 

 

 

 

 

 

 

 

O2 Gas 2

_____%

 

SHEET 1)

 

 

 

 

 

 

 

 

 

 

O2-Reset Vals

Yes/No

 

 

 

 

 

 

 

 

 

 

 

O2 Out Tracks

Yes/No

 

 

 

 

Calib Setup

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

O2 Cal Intervl

____H

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

O2-Next Cal

____H

 

 

 

 

 

 

 

 

 

 

 

 

Gas Time

___Sec.

 

 

 

 

 

 

 

 

 

 

 

 

Purge Time

___Sec.

 

 

 

 

 

 

 

 

 

 

 

 

Auto Calib?

Yes/No

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

O2 Type

_______

 

 

 

 

 

 

 

 

Analog

 

 

O2 Range

______%

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

O2 Alarm Level

_____mA

 

 

 

 

Input/Output

 

 

 

 

 

 

Do O2 Trim

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Logic IO Mode

See Table 4-1

 

 

 

 

 

 

 

 

 

 

 

 

Low O2 Alarm See para. 9-3b

 

 

 

 

 

 

 

 

Digital

 

 

Input State

_______

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Force Output

_______

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

O2 Slope

____ mV/D

 

 

 

 

 

 

 

 

 

 

 

 

O2 Constant

____ mV

 

 

 

 

 

 

 

 

 

 

 

 

O2 T90 Time

0:00

 

SYSTEM

 

 

Parameters

 

 

 

 

 

 

Auto Tune?

Yes/No

 

 

 

 

 

 

 

 

 

Lockout Time

0:00

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Revert Time

0:00

 

 

 

 

 

 

 

 

 

 

 

 

Luminance

______

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Version

xxx

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Checksum

xxx

 

 

 

 

Software

 

 

 

 

 

 

Build Number

xxx

 

 

 

 

 

 

 

 

 

 

Build Date

xxxxxx

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Test Code

xx

 

 

 

 

 

 

 

 

 

 

 

 

SW Err File

xx

 

 

 

 

 

 

 

 

 

 

 

 

SW Err Line

xx

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Status

 

 

 

 

 

 

Alarms

__________

 

 

 

 

 

 

 

 

 

 

PID Parameters

115/220

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Reset Device?

Yes/No

 

 

 

 

 

 

 

 

 

 

 

 

(Cal. required after reset)

 

 

 

 

 

 

 

 

 

 

 

 

 

 

NOTE

In column four of this menu, the selections in Italics are user configurable. Bold text selections are procedures; related instructions are displayed on the LOI. All other parameters are display only.

37260018

Figure 6-4. Menu Tree for Local Operator Interface (LOI) on the Oxymitter 4000 (Sheet 2 of 2)

6-6 OXYMITTER 4000 SETUP AT THE LOI

In setting up the Oxymitter 4000 from the LOI, it’s best to start at the SYSTEM/Calibration Setup menu, Figure 6-4.

a.SYSTEM/Calibration Setup

1.O2 Gas #1 – Enter the high or low cal gas value (the order is not important).

2.O2 Gas #2 – Enter the second cal gas value.

NOTE

Refer to Section 9, MAINTENANCE AND SERVICE, for calibration instructions.

NOTE

Rosemount Analytical recommends 0.4% O2 and 8% O2 for calibration gases.

3.O2 – Reset Values- resets factory de- fault values.

4.O2 Output Tracks – 4 to 20 mA signal can be held at the last value during calibration, or the signal can be left to track the cal gases.

5.O2 Cal Interval – If automatic calibra- tion is selected, this selects the interval between calibrations.

6-4 Startup and Operation with LOI

Rosemount Analytical Inc. A Division of Emerson Process Management