SYMPTOM — NO TEST GAS FLOW

CHECK ALL WIRING BETWEEN

OXYMITTER 4000 AND SPS 4000.

IS

 

 

 

WIRING

 

 

PROPERLY CONNECT WIRING OR

PROPERLY

NO

SECURE LOOSE WIRING CON-

CONNECTED

 

 

NECTIONS; REPLACE DAMAGED

AND

 

 

 

 

WIRING.

SECURE?

 

 

 

 

 

YES

CHECK LOGIC I/O SETTING VIA

HART/AMS.

IS

LOGIC I/O NO

SET FOR

MODE 8?

YES

DISCONNECT CAL GAS INPUT LINES AT MANIFOLD.

SET LOGIC I/O TO MODE 8 VIA HART/AMS.

IS THERE

NO

REPLACED CLOGGED CAL GAS

 

 

LINE BETWEEN CAL GAS

 

FLOW?

CYLINDER AND MANIFOLD.

 

 

 

 

YES

ENSURE CAL GAS FLOWMETER KNOB IS TURNED COUNTERCLOCKWISE TO ALLOW FLOW.

DOES

CAL GAS NO

FLOWMETER

REGISTER

FLOW?

YES

CHECK FUSE F1 ON POWER SUPPLY BOARD.

ISYES

FUSE

BLOWN?

NO

CONTINUED

ON SHEET

2 OF 2

REPLACED CLOGGED CAL GAS LINE BETWEEN MANIFOLD AND CAL GAS FLOWMETER.

REPLACE FUSE PER

PARAGRAPH 9-5a.

F1

J2

HI GAS

LO GAS

NO GAS CAL RET

POWER SUPPLY BOARD

J3

J4 J5

37260040

Figure 8-19. SPS 4000 Troubleshooting Flowchart (Sheet 1 of 2)

Rosemount Analytical Inc. A Division of Emerson Process Management

Troubleshooting 8-23