Oxymitter 4000

Table 8-3. SPS 4000 Fault Finding
SymptomCheck

Fault

Remedy

 

 

 

 

No calibration

Wiring

Improper wire connections,

Properly connect wiring or secure

gas flow

 

loose connections, or

loose wiring connections; replace

 

 

damaged wiring

damaged wiring if necessary.

 

Logic I/O

Oxymitter 4000 logic I/O

Set logic I/O to mode 8 via

 

 

not set for calibration

HART/AMS.

 

 

handshaking with SPS

 

 

 

4000

 

 

Calibration gas lines be-

Clogged calibration gas

Replace clogged calibration gas line.

 

tween cylinders and mani-

line

 

 

fold

 

 

 

Calibration gas flowmeter

Flowmeter knob not turned

Turn calibration gas flowmeter knob

 

knob

counterclockwise to allow

counterclockwise to allow calibration

 

 

flow

gas to flow.

 

 

 

 

 

Calibration gas line be-

Clogged calibration gas

Replace clogged calibration gas line.

 

tween manifold and cali-

line

 

 

bration gas flowmeter

 

 

 

Fuse on power supply

Blown fuse

Replace fuse per paragraph 9-5a.

 

board

 

 

 

Interface board operation

Interface board not send-

Replace interface board per para-

 

 

ing signals

graph 9-5b.

 

Check valve

Clogged check valve

Replace check valve per paragraph

 

 

 

9-5e.

 

Calibration gas line be-

Clogged calibration gas

Replace calibration gas line.

 

tween calibration gas

line

 

 

flowmeter and check valve

 

 

 

Calibration gas flowmeter

Clogged flowmeter

Replace flowmeter per paragraph

 

 

 

9-5h.

 

Power supply output

Power supply failure

Replace power supply board per

 

 

 

paragraph 9-5b.

 

Solenoid

Solenoid failure

Replace solenoid per paragraph

 

 

 

9-5c.

 

Pressure switch

Pressure switch failure

Replace pressure switch per para-

 

 

 

graph 9-5d.