1-7 SPECIFICATIONS Oxymitter 4000

Oxymitter 4000

O2 Range:

 

 

Standard ..................................................

0 to 10% O2, 0 to 25% O2, 0 to 40% O2 (via HART)

Accuracy ..................................................

±0.75% of reading or 0.05% O2, whichever is greater

System Response to Calibration Gas ......

Initial – less than 3 seconds, T90 – less than 8 seconds

Temperature Limits:

 

 

Process ....................................................

32° to 1300°F (0° to 704°C) up to 2400°F (1300°C) with

 

optional accessories

 

Electronics ...............................................

-40° to 185°F (-40° to 85°C) [Operating temperature of

 

electronics inside of instrument housing, as measured by a

 

HART communicator or Rosemount Asset Management

 

Solutions software.]

 

Local Operator Interface ..........................

-40° to 185°F (-40° to 85°C) [Above 158°F (70°C), the in-

 

frared keypad will cease to function, but the Oxymitter

 

4000 will continue to operate properly.]

Probe Lengths ...............................................

18 in. (457 mm)

12 ft (3.66 m)

 

3 ft (0.91 m)

15 ft (4.57 m)

 

6 ft (1.83 m)

18 ft (5.49 m)

 

9 ft (2.74 m)

 

Mounting and Mounting Position ...................

Vertical or horizontal

 

 

Spool pieces, P/N 3D39761G02, are available to offset

Materials:

transmitter housing from hot ductwork.

 

 

Probe .......................................................

Wetted or welded parts - 316L stainless steel (SS)

 

Non-wetted parts - 304 SS, low-copper aluminum

Electronics Enclosure ..............................

Low-copper aluminum

 

Calibration .....................................................

Manual, semi-automatic, or automatic

Calibration Gas Mixtures Recommended .....

0.4% O2, Balance N2

 

 

8% O2, Balance N2

 

Calibration Gas Flow .....................................

5 scfh (2.5 l/m)

 

Reference Air ................................................

2 scfh (1 l/m), clean, dry, instrument-quality air

 

(20.95% O2), regulated to 5 psi (34 kPa)

Electronics .....................................................

NEMA 4X, IP66 with fitting and pipe on reference exhaust

 

port to clear dry atmosphere

 

Electrical Noise ..............................................

Meets EN 55022 Generic Emissions Std.,

 

Includes EN 61000-4-2 for Electrostatic Discharge

 

Includes EN 61000-4-3 for RFI

 

 

Includes EN 61000-4-4 for Fast Transients

 

Includes EN 61000-4-6 for RFI

 

Line Voltage ..................................................

90-250 VAC, 48/62 Hz. No configuration necessary. 3/4 in.

Signals:

- 14 NPT conduit port

 

 

 

Analog Output/HART ...............................

4-20 mA isolated from power supply, 950 ohms max. load

Logic I/O ..................................................

Two-terminal logic contact configurable as either an alarm

 

output or as a bi-directional calibration handshake signal to

 

IMPS 4000 or SPS 4000

 

1-14 Description and Specifications