Instruction Manual

IB-106-340 Rev. 3.0

December 2003

 

HEATER T/C

 

DIAGNOSTIC

HEATER

SW2

ALARMS

O2 CELL

ON

 

 

CALIBRATION

 

CALIBRATION RECOMMENDED

Oxymitter 4000

h. Fault 8, Low Heater Temp

Figure 8-10shows the electronic assembly

for an Oxymitter 4000 with a membrane

keypad (upper view) and an Oxymitter 4000

with an LOI (lower view).

Membrane Keypad. When Fault 8 is de-

tected, the HEATER LED flashes four times,

TEST

POINTS

INC INC

HIGH LOW

GAS GAS

DEC DEC

O2 CELL mV +

O2 CELL mv - HEATER T/C + HEATER T/C -

CAL

TEST GAS + PROCESS - % O2

TP1 TP2 TP3 TP4

TP5 TP6

J1 RED YEL GRN ORG

pauses for three seconds, and repeats.

1.

The low heater temperature alarm is

 

active when the thermocouple reading

 

has dropped below 28.6 mV.

2.

If the thermocouple reading continues

 

to ramp downward for one minute and

 

does not return to the temperature set

 

point of approximately 29.3 mV, then

 

an Open Heater fault will be displayed.

3.

Power down the electronics. Remove

KEYPAD

Alarms

O2 Temp Low

LOI

37260026

Figure 8-10. Fault 8, Low Heater Temp

the electronic assembly per paragraph

9-4d,Electronic Assembly Replace-

ment. Using a multimeter, measure

across the heater connector, J8.

4. If the heater is good, the reading will be

approximately 70 ohms. If the heater is

open, see paragraph 9-4h,Heater Strut

Replacement.

LOI. When Fault 8 is detected, the LOI displays the “O2 Temp Low” message. Refer to the comments and procedures in paragraphs 1 through 4 above.

8-12 Troubleshooting

Rosemount Analytical Inc. A Division of Emerson Process Management