SECTION 3

CONFIGURATION OF OXYMITTER 4000

WITH MEMBRANE KEYPAD

Install all protective equipment covers and safety ground leads before equipment startup. Failure to install covers and ground leads could result in serious injury or death.

3-1 GENERALa.Verify Mechanical Installation

Ensure the Oxymitter 4000 is installed cor- rectly (Section 2, INSTALLATION).

b.Verify Terminal Block Wiring

1.Remove screw (32, Figure 9-3or

Figure 9-4), gasket (33), and cover lock

(34)that secure the housing cover

(27). Remove the cover to expose the terminal block (25).

2.Check the terminal block wiring, Figure 3-1.Be sure the power, 4-20 mA sig- nal, and logic outputs are properly connected and secure.

3.Install the housing cover (27, Figure

9-3or Figure 9-4)on the terminal block (25) and secure with cover lock (34), gasket (33), and screw (32).

4.For an Oxymitter 4000 with an inte- grally mounted SPS 4000, remove screws (26, Figure 9-13)and terminal cover (27). Check that the power and signal terminations are properly con- nected to terminal strip (25) and secure according to instructions in Section 2,

INSTALLATION.

5.Install terminal cover (27) and secure with screws (26). Make sure terminal cover gasket (28) is in place.

+ 20-4-

-

+

 

 

 

AC

AC

 

 

N

L1

4-20 mA

 

SIGNAL

TERMINAL

 

BLOCK

LOGIC I/O

 

GROUND LUGS

OXYMITTER 4000

ELECTRONICS

HOUSING

Rosemount Analytical Inc.

SMART FAMILY

Orrville, OH 44667-0901

800-433-6076

HART

OXYMITTER 4000

SERIAL NO.

TAG NO.

VOLTS: 85-264 VAC

48-62 Hz WATTS:

500 VA

OUTPUT: 4-20 mA

LINE FUSE:

5 Amps

HEATER T/C

DIAGNOSTIC HEATER SW2

ALARMS 02 CELL

CALIBRATION

CALIBRATION RECOMMENDED

 

 

02 CELL mV +

TP1

J1

TEST

02 CELL mv -

TP2

 

POINTS

HEATER T/C +

TP3

 

 

 

HEATER T/C -

TP4

RED YEL GRN ORG

 

 

 

 

INC

INC

CAL

 

 

 

 

 

 

HIGH

LOW

 

 

 

GAS

GAS

 

 

 

DEC

DEC

TEST GAS +

TP5

 

PROCESS -

TP6

 

 

 

 

 

 

% 02

 

 

37260009

Figure 3-1. Electronics Housing Terminals and Membrane Keypad

Rosemount Analytical Inc. A Division of Emerson Process Management

Configuration with Keypad 3-1