Instruction Manual

IB-106-340 Rev. 3.0 December 2003

Oxymitter 4000

 

0.125-27 NPT FEMALE

3.12 (79.25) MAX

 

 

 

OUTLET CONNECTION

 

 

NOTE: DIMENSIONS ARE IN INCHES WITH

 

1

 

 

2.250 (57.15)

 

MILLIMETERS IN PARENTHESES.

 

2

OUTLET

3

 

 

 

 

 

 

 

 

 

 

 

4.81

(122.17)

 

 

 

0.25-18 NPT FEMALE

 

 

 

 

 

 

 

 

 

 

 

 

INLET CONNECTION

8.50

 

 

 

 

 

 

 

 

(215.90)

FLOW SET

 

 

 

 

 

 

 

MAX

POINT KNOB

 

 

1.19

2.0

 

 

 

 

 

 

 

 

2 MOUNTING HOLES

 

 

 

(30.22)

 

 

 

 

(50.80)

 

 

 

 

3.19

(81.03)

LG

 

 

 

 

 

 

1.50THROUGH BODY FOR

 

DRAIN VALVE

(38.10)

0.312 (7.92) DIA BOLTS

 

 

 

 

 

 

 

 

10.0

 

 

 

 

 

 

(254)

 

 

 

 

 

 

REF

 

 

 

 

 

 

 

1

FLOWMETER

 

0.2-2.0 SCFH

771B635H02

 

 

2

2" PRESSURE GAGE

0-15 PSIG

275431-006

 

TO ELECTRONICS

3

COMBINATION FILTER-REG.

0-30 PSIG

4505C21G01

 

 

 

 

 

 

OXYMITTER 4000

 

 

 

 

 

 

SERIAL NO.

 

 

 

 

 

 

TAG NO.

 

 

 

 

 

 

VOLTS: 85-264 VAC 48-62 Hz

WATTS: 500 VA

 

 

 

 

 

OUTPUT: 4-20 mA LINE FUSE:

5 Amps

 

 

 

 

 

 

 

 

 

INSTRUMENT AIR SUPPLY

 

 

 

 

 

10-225 PSIG MAX PRESSURE

REF AIR SET

0.250 OR 6 MM O.D. TUBING263C152G01

(SUPPLIED BY CUSTOMER)

SCHEMATIC HOOKUP FOR REFERENCE AIR SUPPLY ON OXYMITTER 4000 PROBE HEAD.

26170035

Figure 2-13. Air Set, Plant Air Connection

 

SMART FAMILY

 

HART

OXYMITTER 4000

 

SERIAL NO.

 

TAG NO.

 

VOLTS: 85-264 VAC WATTS:48-62 Hz

500 VA

OUTPUT: 4-20 mALINE FUSE:

5 Amps

 

REFERENCE AIR

CALIBRATION GAS

26170025

Figure 2-14. Oxymitter 4000 Gas Connections Calibration Gas Connections

2-18 Installation