MODEL 54e pH/ORP

SECTION 5.0

 

SOFTWARE CONFIGURATION

TABLE 5-1. pH Settings List (continued)

ITEM

 

RANGES

FACTORY SETTINGS

USER SETTINGS

5.

Alarm 3 Control

 

 

 

 

 

 

(a) Activation Method

 

Process/Temp

 

Process

_______

 

(b) Alarm 3 Control Mode

 

Normal/TPC

 

Normal

_______

6a. Alarm 3 Setup (Normal)

 

 

 

 

 

 

(a) Configuration

 

Low alarm/High alarm/Off

 

High

_______

 

(b) Hysteresis (deadband)

 

0 - 5.00 pH

 

0.01 pH

_______

 

(e) Delay Time

 

0-99 sec

 

0 sec

_______

 

(d) Relay Fault

 

Open/Closed/None

 

None

_______

6b. Alarm 3 Setup (TPC)

 

 

 

 

 

 

(a) Setpoint

-2 to 16 pH or -15 to 130°C

 

7 pH

_______

 

(b) Proportional

 

0-299.9%

 

100.0%

_______

 

(c) Integral

 

0-2999 sec

 

0 sec

_______

 

(d) Derivative

 

0-299.9%

 

0.0%

_______

 

(e) Time Period

 

10-2999 sec

 

30 sec

_______

 

(f) LRV (100% On)

-2 to 16 pH or -15 to 130°C

 

2 pH

_______

 

(g) URV (100% Off)

-2 to 16 pH or -15 to 130°C

 

0 pH

_______

 

(h) Relay Fault

 

None/Open/Closed

 

None

_______

7.

Alarm 4 Control

 

 

 

 

 

 

(a) Alarm

 

Fault/Off

 

Fault

_______

8.

Feed Limit Timer

 

 

 

 

 

 

(a) Feed Limit

Disable/alarm 1/alarm 2/alarm 3

Disable

_______

 

(b) Timeout Value

 

0-10,800 sec

 

3600 sec

_______

C. Alarms (Section 5.7)

 

 

 

 

 

9.

Interval Timer

 

 

 

 

 

 

(a) Timer (selection)

Disable/alarm 1/alarm 2/alarm 3

Disable

_______

 

(b) Timer (activation method)

Time activated/Imped. activated

Time activated

_______

 

(c) Interval

 

0-999.9 hr

 

24.0 hr

_______

 

(d) Repeats

 

1-60

 

1

_______

 

(e) On Time

 

0-2999 sec

 

120 sec

_______

 

(f) Off Time

 

0-2999 sec

 

1 sec

_______

 

(g) Recovery

 

0-999 sec

 

600 sec

_______

D. Diagnostics (Section 5.8)

 

 

 

 

 

1.

Diagnostics (Glass and Reference)

 

On/Off

 

000 (no security)

_______

2.

Glass Imp(edance) High Setpoint

0-2000 MOhms (0 disables)

 

000 (no security)

_______

3.

Glass Imp(edance) Low Setpoint

 

0-900 MOhms (0 disables)

 

 

 

4.

Ref(erence) Imp(edance) High

 

0-140 KOhms (0 disables)

 

000 (no security)

_______

5.

Zero Offset

 

0-999 mV (0 disables)

 

000 (no security)

_______

6.

Cal(ibration) Warn(ing)

 

0-500% (0 disables)

 

 

 

7.

Imped(ance Temperature) Comp(ensation)

On/Off

 

000 (no security)

_______

E. Auto Calibration (Section 5.9)

 

 

 

 

 

1.

Autocal (Buffer List Selection)

Manual/Standard/DIN 19267

 

000 (no security)

_______

2.

Stabilize pH (Auto Calibration)

 

.01-.50 pH

 

000 (no security)

_______

3.

Stabilization Time (Auto Calibration)

 

0-30 sec

 

000 (no security)

_______

F. Security (Section 7.1)

 

 

 

 

 

1.

Lock all

 

000-999

 

000 (no security)

_______

2.

Lock Program (Lock all except Calibrate)

000-999

 

000 (no security)

_______

3.

Lock Config. (Lock all except Calibrate,

 

 

 

 

 

 

Output setpoints (PID), Simulated Tests

 

 

 

 

 

 

Alarm Setpoints, and Rerange Outputs)

 

000-999

 

000 (no security)

_______

D. Solution Temperature Compensation (Section 7.3)

 

 

 

1. Temp(erature) Coeff(icient)

 

-0.044 to 0.028pH/°C

 

000 (no security)

_______

2.

Operate Iso(potential pH)

 

-1.35 to 20.12 pH

 

000 (no security)

_______

3.

Sensor Iso(potential)

 

0-14 pH

 

000 (no security)

_______

 

 

 

 

 

 

 

By changing the standard output configuration, you can set up the Model 54e pH/ORP to perform a wide variety of control and monitoring tasks. The configuration procedures allow you to program the controller to meet the specific control and monitoring requirements of your particular plant. This is done by recording the desired configuration parameters on the List of Settings Form and then actually configuring them by using the keys on the controller front panel.

24

Page 30
Image 30
Emerson 54e pH/ORP instruction manual Diagnostics Section