MODEL 54e pH/ORP

SECTION 1.0

 

DESCRIPTION AND SPECIFICATIONS

The Model 54e when configured as a pH analyzer, requires a dual (glass and reference) impedance pre- amplifier (patent pending). This preamp converts the high impedance pH glass electrode signal to a low impedance signal. The preamplifier may be located in one of three areas: 1) in the pH sensor for best per- formance, 2) in a remote junction box when process temperatures exceed 80°C (176°F) in submersion applications, or 3) in the analyzer when the distance between the pH sensor and the analyzer is 4.5 meters (15 feet) or less.

The Model 54e pH measures over the full range of 0- 14 pH. The current output may be calibrated to repre- sent any 1 to 14 pH span.

A two-point calibration is made by immersing the sen- sor in two different buffer solutions and entering the pH values. When two buffers are used, the micro- processor automatically calculates the electrode slope which is used for self-diagnostics. The electrode slope can be read on the display and manually adjusted. A one-point process standardization is easily accom- plished by entering the pH value of a grab sample.

ANALYZER SPECIFICATIONS @ 25°C

Measurement Range: 0 to 14 pH

Output Scale Expansion: Zero suppression: up to 13 pH units

Span: Any pH from 1 to 14

Accuracy: ± 0.01 pH

Repeatability: ± 0.01 pH

Stability: ± 0.01 pH/month, non-cumulative

Temperature Coefficient: Input: ± 0.003 pH/°C

Output: ± 0.006 pH/°C

Temperature Compensation : Pt 100 or Pt 1000 RTD, Automatic or Manual

–15 to 120°C (5 to 248°F)

The Model 54e when configured as an ORP analyzer, measures over a range of -1400 mV to +1400 mV in either the American convention (Oxidation Reduction Potential), or the European con- vention (Reduction Oxidation-Redox). Although tem- perature compensation is not used for ORP measurements, the process temperature is measured and displayed. Temperature measurement is made by an RTD located in the sensor assembly.

ANALYZER SPECIFICATIONS @ 25°C

Measurement Range: –1400 to +1400 mV

Output Scale Expansion: Zero suppression: up to ±1300 mV

Span: Any ORP range from 100 to 2800 mV

Accuracy: ± 1.0 mV

Repeatability: ± 1.0 mV

Stability: ± 1.0 mV/month, non-cumulative

Temperature Coefficient: Input: ± 0.2 mV/°C

Output: ± 0.4 mV/°C

Temperature Measurement: –15 to 120°C (5 to 248°F) Pt 100 or Pt 1000 RTD

RECOMMENDED SENSORS:

Model 330 Flow Through ORP

*Model 371 EuroSenz ORP

Model 381+ Insertion/Submersion/Flow Through ORP *Model 389 Disposable ORP

Model 396P Disposable ORP

Model 396R Retractable ORP

Model 398 Insertion/Submersion ORP

Model 398VP Insertion/Submersion with VP 6.0 connector Model 398R Retractable ORP

Model 398RVP Retractable ORP with VP 6.0 connector

RECOMMENDED SENSORS:

Model 320B Flow Through pH

Model 320HP High Purity pH *Model 328A Steam Sterilizable pH *Model 370 and 371 EuroSenz pH

Model 381+ Insertion/Submersion/Flow Through pH

*Model 389 Disposable pH *Model 396/VP Disposable pH Model 396P/VP Disposable pH Model 396R/VP Retractable pH Model 397 Quik Disconnect pH

Model 398/VP Insertion/Submersion pH

*Model 398R/VP Retractable pH *Model 399 Disposable pH

*Model Hx338 Steam Sterilizable pH

*Model Hx348 Steam Sterilizable pH

The Model 54e when ordered for ISE capability,

is suitable for use with a number of ion-selective elec- trodes. Consult the factory for available measure- ments and ranges.

3

Page 9
Image 9
Emerson 54e pH/ORP instruction manual Analyzer Specifications @ 25C, Recommended Sensors