MODEL 54e pH/ORP |
| SECTION 8.0 |
|
| TROUBLESHOOTING |
TABLE |
| |
|
|
|
PROBLEM OR CONDITION | PROBABLE CAUSE AND | FOR MORE HELP, |
| CORRECTIVE ACTION | REFER TO |
Controller "short spans." Buffer check with
"lo slope error"
Change in pH slope (value shrinking) after
pH value locks up (no change in buffers).
Coated glass electrode - Clean with soft cloth and clean water.
Service 381+ Reference Junction
and Solution.
Old glass electrode -
If used for more than 1 yr, replace electrode.
Old or coated glass electrode -
Clean or replace glass electrode. |
|
Bad RTD value - | Temp.Comp. Troubleshooting |
See Temp. Comp. Troubleshooting | Procedure, Section 8.2 |
Procedure with expected RTD value |
|
resistances at temperature values. |
|
Glass electrode cracked - |
|
Replace glass electrode. |
|
Coated glass electrode - |
|
Clean glass electrode. |
|
Displayed pH value not the same | Grab sample incorrect - | Calibration, Section 4.0 |
as grab sample of process. | Reevaluate sample technique and |
|
| equipment. |
|
| Unclear which is correct - | Calibration, Section 4.0 |
| Retest with new buffer solution. |
|
| Controller out of calibration - | Calibration, Section 4.0 |
| Recalibrate per Calibration, |
|
| (See Section 3.0) |
|
Display reads off scale. | (381+ Sensor) Reference depleted - |
|
| Replace reference electrolyte as |
|
| instructed in 381+ manual. |
|
| Defective preamplifier - | Section 8.0 |
| Check preamplifier as instructed in |
|
| Preamp Troubleshooting Procedure, and |
|
| replace preamplifier, if defective. |
|
| Electrode not in process solution - | Wiring, Section 3.0 |
| Make sure sensor is in solution. |
|
| Open circuit within glass electrode - |
|
| Replace glass electrode. |
|
| (381+ Sensor) Plugged reference - |
|
| Replace reference junction. |
|
| Recharge reference solution. |
|
Table
67