MODEL 54e pH/ORP

 

SECTION 8.0

 

 

TROUBLESHOOTING

TABLE 8-3. Troubleshooting Guide (continued)

 

 

 

 

PROBLEM OR CONDITION

PROBABLE CAUSE AND

FOR MORE HELP,

 

CORRECTIVE ACTION

REFER TO

Controller "short spans." Buffer check with 2-point calibration gives slope less than 47 mV/pH .

"lo slope error"

Change in pH slope (value shrinking) after 2-point calibration.

pH value locks up (no change in buffers).

Coated glass electrode - Clean with soft cloth and clean water.

Service 381+ Reference Junction

and Solution.

Old glass electrode -

If used for more than 1 yr, replace electrode.

Old or coated glass electrode -

Clean or replace glass electrode.

 

Bad RTD value -

Temp.Comp. Troubleshooting

See Temp. Comp. Troubleshooting

Procedure, Section 8.2

Procedure with expected RTD value

 

resistances at temperature values.

 

Glass electrode cracked -

 

Replace glass electrode.

 

Coated glass electrode -

 

Clean glass electrode.

 

Displayed pH value not the same

Grab sample incorrect -

Calibration, Section 4.0

as grab sample of process.

Reevaluate sample technique and

 

 

equipment.

 

 

Unclear which is correct -

Calibration, Section 4.0

 

Retest with new buffer solution.

 

 

Controller out of calibration -

Calibration, Section 4.0

 

Recalibrate per Calibration,

 

 

(See Section 3.0)

 

Display reads off scale.

(381+ Sensor) Reference depleted -

 

 

Replace reference electrolyte as

 

 

instructed in 381+ manual.

 

 

Defective preamplifier -

Section 8.0

 

Check preamplifier as instructed in

 

 

Preamp Troubleshooting Procedure, and

 

 

replace preamplifier, if defective.

 

 

Electrode not in process solution -

Wiring, Section 3.0

 

Make sure sensor is in solution.

 

 

Open circuit within glass electrode -

 

 

Replace glass electrode.

 

 

(381+ Sensor) Plugged reference -

 

 

Replace reference junction.

 

 

Recharge reference solution.

 

Table 8-3 is continued on the following page

67

Page 73
Image 73
Emerson 54e pH/ORP instruction manual Coated glass electrode Clean with soft cloth and clean water