MODEL 54e pH/ORP

 

SECTION 8.0

 

 

TROUBLESHOOTING

TABLE 8-3. Troubleshooting Guide

 

 

 

 

PROBLEM OR CONDITION

PROBABLE CAUSE AND

FOR MORE HELP,

 

CORRECTIVE ACTION

REFER TO

Controller completely inoperative

No Power -

Wiring, Section 3.0 and Figure 3-1

 

Check power supply at breaker and

 

 

inside controller: 115 V across

 

 

terminals 1 and 2 on TB3.

 

 

Cut-off circuit -

 

 

Check power supply board.

Figure 3-1

Electronics Failure -

Replace the electronics.

Controller operating, but adding chemical above setpoint, or not adding below setpoint,

or not holding setpoint.

Incorrect or Changed Settings -

Software Configuration,

Refer to software configuration

Section 5.0

procedure and verify that control

 

parameters are correct and entered

 

properly. Pay special attention

 

to the TPC settings.

 

Electronics Failure -

Simulated tests

Try power down and power back

Section 5.4

up to reboot the program. Test

 

alarm relays.

 

Replace electronics if necessary.

 

Inability to standardize pH

Incorrect standardization procedure -

Calibration, Section 4.0

"Zero Offset Err"

If you are trying to adjust the pH reading

 

 

by a large amount, the controller may

 

 

reject your standardization and will

 

 

continue to display the previous value.

 

 

Check that your lab instrument, titrations,

 

 

or chemical standard solutions have

 

 

been used properly and are correct.

 

 

Sensor failure -

Section 8.2

 

Simulate sensor in "Preamp

 

 

Troubleshooting".

 

 

Replace sensor if preamp ok.

 

 

Electrode coated -

 

 

Clean electrode as needed.

 

 

(381+ Sensor) Reference chamber

 

 

is contaminated -

 

 

Replace gel-filled solution as instructed

 

 

in 381+ manual

 

 

Defective preamplifier -

Section 8.0

 

Check preamplifier as instructed in

 

 

Preamp Troubleshooting procedure

 

 

and replace if defective.

 

 

(381+ Sensor) Bad Reference -

 

 

If trouble still persists, microjunction on

 

 

reference electrode is defective.

 

 

Replace reference electrode.

 

66

Table 8-3 is continued on the following page

 

Page 72
Image 72
Emerson 54e pH/ORP instruction manual Troubleshooting Guide, No Power