MODEL 54e pH/ORP

 

SECTION 8.0

 

 

TROUBLESHOOTING

TABLE 8-3. Troubleshooting Guide (continued)

 

 

 

 

PROBLEM OR CONDITION

PROBABLE CAUSE AND

FOR MORE HELP,

 

CORRECTIVE ACTION

REFER TO

Display reads between 3 and

Electrode cracked -

 

6 pH regardless of actual pH

Replace electrode.

 

solution or sample.

 

 

"cracked glass failure"

Sluggish or slow display indica- tion for real changes in pH level.

Inability to change parameters in the controller. "Level 1 security: Lock" "Level 2 security: Lock" "Level 3 security: Lock"

Outputs do not change "Hold Mode activated" "simulating output 1 or 2" "simulating alarm 1,2,3, or 4"

Red LED light is on Various fault messages shown on lower display.

Electrode coated -

Clean electrode as instructed in sensor manual

Password protected -

Section 7.0

Your controller has password protection.

 

You must enter the correct password to

 

make changes in the controller.

 

Controller is in Hold or simulate mode -

Software Configuration, Section 5.6

To remove from Hold, press any key

Calibration , Section 4.7

and then press Enter (F4). Press Edit (F4),

 

change "On" to Off", and press Save (F4).

 

To stop simulation, press Exit (F1).

 

When the unit has been put into "Hold",

 

all outputs go to their default states.

 

Unit has gone into fault mode -

Section 8.0

Read the message and take action as

 

indicated in Table 8-2.

 