MODEL 54e pH/ORP

 

 

 

SECTION 7.0

 

 

 

 

 

 

SPECIAL PROCEDURES AND FEATURES

7.3 SOLUTION TEMPERATURE COMPENSATION (continued)

 

 

 

 

 

 

 

 

 

 

 

2.

 

Menu Item

Options

 

 

Temp coeff: +.000

 

 

 

 

 

 

 

 

 

 

Temperature coefficient

-0.044 to 0.028 pH/°C (0.00 for

 

 

Operate iso: +07.00 pH

 

most applications)

 

 

 

 

 

Operating isopotential

-1.35 to 20.12 pH (7.00 for most

 

 

Sensor iso: 07.00 pH

 

 

(Isopotential pH)

applications)

 

 

 

 

 

 

 

 

 

 

Exit

Edit

Sensor isopotential

0-14 pH (7.00 for most applications)

 

 

 

 

 

 

 

 

 

 

 

 

 

The values now being used by the controller are displayed. To change any of these items, use the arrow key to highlight the desired item and press Edit 4. Use the arrow keys to make the change and press Save 4 to enter the change into memory.