MODEL 54e pH/ORP

SECTION 4.0

 

CALIBRATION

4.7 HOLD MODE

7.00pH

Hold Mode: Off

Placing the Controller on Hold for Maintenance. Before performing main- tenance or repair of the sensor, the Controller can be placed in hold (refer to Section 5.6 to enable this feature) to prevent process upsets while the reading is off-line. This will place the current outputs into the selected default states (see Section 5.6). The relays will act as selected in relay default, see Section 5.7.

Exit

Cont

Edit

Before removing the sensor from the process, press any key and then

 

 

 

 

 

 

Enter (F4). When the hold mode has been enabled, the hold mode

 

 

 

screen (on the left) will appear prior to calibration. To continue without

 

 

 

putting the controller in hold, simply press Cont (F3). To put the controller

 

7.00 pH

 

in hold, press Edit (F4), use the arrow key to change the "Off" to "On" and

 

 

press Save (F4).

 

 

 

 

 

 

NOTE

26.2°C 12.0mA

Hold Mode Activated

When the Hold Mode is activated ("On"), "Hold Mode Activated" will always appear on the bottom line of the display.

Always calibrate after cleaning or repair of the pH sensor. After installing the sensor back into the process, always change the Hold Mode setting to OFF.

4.8 TRIM OUTPUTS

Temp compensation pH slope

Output trim

ExitEnter

The instrument’s current outputs may be calibrated (trimmed) if neces- sary. If either the power board or the CPU board is replaced, the outputs must be calibrated. To perform this procedure, a calibrated meter must be connected to the output being calibrated.

To perform an output calibration, from the main display press any key to obtain the main menu. With the cursor on “calibrate,” press Enter (F4). With the cursor on “Output trim,” press Enter (F4) again. Select “Trim out- put 1” or “Trim output 2” as appropriate.

Press Edit (F4) to select Cal point 1 (4 mA expected and simulated) or Cal point 2 (20 mA expected and simulated). Adjust the Meter value to match the reading of the calibrated meter connected to the output. Press Enter (F4) to complete the calibration.

21

Page 27
Image 27
Emerson 54e pH/ORP instruction manual Output trim, Hold Mode, Trim Outputs