MODEL 54e pH/ORP

 

 

 

 

SECTION 5.0

 

 

 

 

 

 

 

SOFTWARE CONFIGURATION

5.7 CHANGING ALARM PARAMETERS (continued)

 

 

 

 

 

 

 

 

 

 

Alarm Control Parameters

3.

 

Menu Item

Options

 

 

 

 

 

 

 

Activation Method

Process/Temperature

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Control Mode

Normal/TPC

 

 

Activation Method

 

 

 

 

 

 

 

 

 

 

 

 

Control Mode

 

 

Alarms 1, 2, and 3 can each be configured with the options above.

 

 

 

 

 

 

 

 

 

 

The default options are that all three alarms are Process (pH or

 

 

 

 

 

ORP/Redox), and Normal (not TPC). This is a common configura-

 

 

 

 

 

tion and may not require changes. If no changes are desired, skip to

 

Exit

Enter

 

step 4a.

 

 

 

 

 

 

 

To make changes in these parameters, highlight the desired menu

 

 

 

 

 

header and press Enter 4. The value now being used is displayed

 

Activate: Process

 

 

 

 

 

 

 

and the F4 key can now be pressed to Edit the item. Once Edit has

 

 

 

 

 

been pressed, change the item as needed and then press Save 4 to

 

 

 

 

 

store the value. Repeat for the other output and/or items as needed.

 

Exit

Edit

 

 

 

 

 

Alarm Setup Parameters (for Normal Alarms only)

Alarm: Low

Setpoint: 0.01 pH

Delay: 0 sec

ExitEnter

Hysteresis: 0.01 pH

Delay: 0 sec

Relay Default: None

ExitEnter

4a.

 

Menu Item

Options

 

 

Alarm (action)

Low/High/Off

 

 

Setpoint

-1400 to1400 mV, -2.00 to 20.00 pH, 0 to 200°C.

 

 

Hysteresis

-2.00 to 20.00 pH, 0 to 10°C.

 

 

 

 

 

 

Delay

0-99 sec

 

 

Relay default

None/Close/Open

 

 

 

 

These parameters can be adjusted by highlighting the desired item and pressing the Edit 4 key. Once Edit has been pressed, change the item as needed and then press Save 4 to store the value.

"Alarm action" determines whether alarm will activate when the reading exceeds the setpoint (high alarm) or when it drops below the setpoint (low action). It can also be turned off (i.e. not used).

"Hysteresis" is a deadband that prevents deactivating a relay until the reading has dropped below the setpoint minus the hysteresis amount (high alarm example).

"Delay" will delay activation (and deactivation) of the relay for a cer- tain number of seconds. Larger delays can reduce relay chatter.

"Relay Default" determines how the relay will act if there is a fault or hold condition. Each alarm can be forced on (Close), off (Open) or can remain unchanged (None). The factory configuration is "None".

37

Page 43
Image 43
Emerson 54e pH/ORP Alarm Low, Setpoint 0.01 pH Delay 0 sec, Hysteresis 0.01 pH Delay 0 sec, Relay Default None