MODEL 54e pH/ORP

 

APPENDIX A

 

 

ORP CONFIGURATION

 

TABLE A-1. ORP Settings List

 

 

 

ITEM

CHOICES

ORP SETTINGS USER SETTINGS

2.Alarm 1 Setup

 

(a) Configuration

Low alarm/High alarm/Off

Low

_______

 

(b) Hysteresis (deadband)

0-500 mV

0 mV

_______

 

(c) Delay Time

0-99 sec

0 sec

_______

 

(d) Relay Default

Open/Closed/None

None

_______

3.

Alarm 2 Control

 

 

 

 

(a) Activation Method

Process/Temp

Process(ORP)

_______

 

(b) Alarm 2 Control Mode

Normal/TPC

Normal

_______

4.

Alarm 2 Setup

 

 

 

 

(a) Configuration

Low alarm/High alarm/Off

High

_______

 

(b) Hysteresis (deadband)

0-500 mV

0 mV

_______

 

(c) Delay Time

0-99 sec

0 sec

_______

 

(d) Relay Default

Open/Closed/None

None

_______

5.

Alarm 3 Control

 

 

 

 

(a) Activation Method

Process/Temp

Process(ORP)

_______

 

(b) Alarm 3 Control Mode

Normal/TPC

Normal

_______

6.

Alarm 3 Setup

 

 

 

 

(a) Configuration

Low alarm/High alarm/Off

High

_______

 

(b) Hysteresis (deadband)

0-500 mV

0 mV

_______

 

(c) Delay Time

0-99 sec

0 sec

_______

 

(d) Relay Default

Open/Closed/None

None

_______

7.

Alarm 4 Control

 

 

 

 

(a) Alarm

Fault/Off

Fault

_______

8.

Feed Limit Timer

 

 

 

 

(a) Feed Limit

Disable/alarm 1/alarm 2/alarm 3

Disable

_______

 

(b) Timeout Value

0-10,800 sec

3600 sec

_______

9.

Interval Timer

 

 

 

 

(a) Timer (selection)

Disable/alarm 1/alarm 2/alarm 3

Disable

_______

 

(b) Timer (activation method)

Time activated/Imped.activated

Time activated

_______

 

(c) Interval

0-999.9 hr

24.0 hr

_______

 

(d) Repeats

1-60

1

_______

 

(e) On Time

0-2999 sec

120 sec

_______

 

(f) Off Time

0-2999 sec

1 sec

_______

 

(g) Recovery

0-999 sec

600 sec

_______

D.Diagnostics (Section 5.8)

1.

Diagnostics (Glass and Reference)

On/Off

Off

_______

4.

Ref(erence) Imp(edance) High

0-140 KOhms (0 disables)

140 KOhms

_______

5.

Zero Offset

0-999 mV (0 disables)

60 mV

_______

E.Security (Section 7)

1.

Lock all

000-999

000 (no security)

_______

2.

Lock Program (Lock all except Calibrate) 000-999

000 (no security)

_______

3.Lock Config. (Lock all except Calibrate,

Alarm Setpoints and Rerange Outputs) 000-999

000 (no security)

_______

73

Page 79
Image 79