MODEL 54e pH/ORP |
| APPENDIX A |
|
| ORP CONFIGURATION |
| TABLE | |
|
|
|
ITEM | CHOICES | ORP SETTINGS USER SETTINGS |
2.Alarm 1 Setup
| (a) Configuration | Low alarm/High alarm/Off | Low | _______ |
| (b) Hysteresis (deadband) | 0 mV | _______ | |
| (c) Delay Time | 0 sec | _______ | |
| (d) Relay Default | Open/Closed/None | None | _______ |
3. | Alarm 2 Control |
|
|
|
| (a) Activation Method | Process/Temp | Process(ORP) | _______ |
| (b) Alarm 2 Control Mode | Normal/TPC | Normal | _______ |
4. | Alarm 2 Setup |
|
|
|
| (a) Configuration | Low alarm/High alarm/Off | High | _______ |
| (b) Hysteresis (deadband) | 0 mV | _______ | |
| (c) Delay Time | 0 sec | _______ | |
| (d) Relay Default | Open/Closed/None | None | _______ |
5. | Alarm 3 Control |
|
|
|
| (a) Activation Method | Process/Temp | Process(ORP) | _______ |
| (b) Alarm 3 Control Mode | Normal/TPC | Normal | _______ |
6. | Alarm 3 Setup |
|
|
|
| (a) Configuration | Low alarm/High alarm/Off | High | _______ |
| (b) Hysteresis (deadband) | 0 mV | _______ | |
| (c) Delay Time | 0 sec | _______ | |
| (d) Relay Default | Open/Closed/None | None | _______ |
7. | Alarm 4 Control |
|
|
|
| (a) Alarm | Fault/Off | Fault | _______ |
8. | Feed Limit Timer |
|
|
|
| (a) Feed Limit | Disable/alarm 1/alarm 2/alarm 3 | Disable | _______ |
| (b) Timeout Value | 3600 sec | _______ | |
9. | Interval Timer |
|
|
|
| (a) Timer (selection) | Disable/alarm 1/alarm 2/alarm 3 | Disable | _______ |
| (b) Timer (activation method) | Time activated/Imped.activated | Time activated | _______ |
| (c) Interval | 24.0 hr | _______ | |
| (d) Repeats | 1 | _______ | |
| (e) On Time | 120 sec | _______ | |
| (f) Off Time | 1 sec | _______ | |
| (g) Recovery | 600 sec | _______ |
D.Diagnostics (Section 5.8)
1. | Diagnostics (Glass and Reference) | On/Off | Off | _______ | |
4. | Ref(erence) Imp(edance) High | 140 KOhms | _______ | ||
5. | Zero Offset | 60 mV | _______ |
E.Security (Section 7)
1. | Lock all | 000 (no security) | _______ | |
2. | Lock Program (Lock all except Calibrate) | 000 (no security) | _______ |
3.Lock Config. (Lock all except Calibrate,
Alarm Setpoints and Rerange Outputs) | 000 (no security) | _______ |
73