MODEL 54e pH/ORP

 

 

 

SECTION 5.0

 

 

 

 

 

 

 

SOFTWARE CONFIGURATION

5.5 CHOOSING DISPLAY OPTIONS

 

 

 

 

 

 

This section describes the options available for the changing of engineering

 

Display

 

 

units and variables on the main display.

 

 

 

Outputs

1. From the main menu, move the cursor down to "Program" and press

 

 

 

 

 

Alarms

 

Enter 4. From the program menu, move the cursor down using the

 

 

arrow key to highlight "Configure" and press Enter 4.

 

Exit

Enter

The first configuration menu is displayed. With the cursor on "Display"

 

press Enter 4.

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

2.

 

Menu Item

Options

 

 

Measure: pH

 

 

 

 

 

 

 

 

 

 

 

 

Measure

pH/ORP/Redox

 

 

Resolution: 0.01

 

 

 

 

 

 

 

 

 

 

 

Resolution (pH only)

0.01pH/0.1pH

 

 

Temp Units: °C

 

 

 

 

 

 

 

 

Temperature units

°C/°F

 

 

Exit

Enter

The values now being used by the controller are displayed. To change

 

 

 

 

 

any of these items, use the arrow key to highlight the desired item and

 

 

 

 

 

press Edit(F4). Use the arrow keys to make the change and press Save

 

 

 

 

 

(F4) to enter the change into memory.

 

 

WARNING: This change resets analyzer to factory settings

Abort Cont

WARNING

Changing the measurement selection from pH to ORP or ORP to pH will reset all controller settings to factory values. Press Abort (F1) to cancel the change. Press Cont (F3) to change the measurement. Refer to Appendix A for more about ORP measurements.

Output 1: mA

3.

 

 

 

 

 

 

 

Output 2: mA

 

Language: English

 

 

Exit

Edit

Menu Item

Options

 

 

Output 1 units

mA/%

 

 

Output 2 units

mA/%

 

 

Language

English/Français/Español/Deutsch/Italiano

 

 

Display left: AL1

Display right: AL2

Display contrast: 5

ExitEdit

Press the arrow key three times to access this screen. The current out- puts can be displayed as milliamps or as percent of full scale. The default is mA. To change any of these items, use the arrow key to highlight the desired item and press Edit (F4). Use the arrow keys to make the change and press Save (F4) to enter the change into memory.

Further menu items are available by pressing the arrow key repeatedly. When the display is highlighting the item on the third line, the end of the menu has been reached. To back up within the menu, use the up arrow key.

Measure: pH

Resolution: 0.01

Temp Units: °C

ExitEnter

31

Page 37
Image 37
Emerson 54e pH/ORP instruction manual Display right AL2, Measure pH