MODEL 54e pH/ORP

APPENDIX A

 

ORP CONFIGURATION

Accessing Calibrate, Program and Configure Menus. Operating configuration changes are made at the levels shown in Figure A-1. Press any key from the main display to access the main menu (top left). Refer to Section 5 for pH measurements.

Level 1 Calibrate. To access calibration selections from the main menu, with the cursor on "Calibrate" press Enter (F4). ORP standardization and tempera- ture adjustments are made at this level (refer to Section A.1 and Section 3.1 for these procedures).

Level 2 Program. To access the program level from the main menu, place the cursor over "Program" with the down arrow key. Then press Enter (F4). From the program level menu, changes can be made to the alarm setpoints and the output setpoints.

Level 3 Configure. To access the configure level from the main menu place cursor over "Program" and Enter (F4) then place cursor over "Configure" and Enter (F4). This level contains advanced selections, such as alarms and diagnostics.

Calibrate

Diagnostic variables

Program

ExitEnter

PRESS ENTER

Standardize

Adjust temperature

ExitEnter

CALIBRATE MENU SEC- TION

Buffer Calibration 4.2, 4.3

Standardize pH

4.4

Adjust Temperature

4.1

Temp Compensation4.5

pH slope

4.6

PRESS TWO ARROWS b

and then ENTER

Alarm setpoints

Output setpoints

Simulated tests

ExitEnter

PRESS

ARROWS

b

 

Simulated tests

 

 

 

 

 

Configure

 

 

 

 

 

 

Exit

Enter

 

 

 

 

PRESS

 

 

 

 

 

PROGRAM MENU

SEC

 

 

ENTER

TION

 

 

 

 

 

 

Alarm setpoints

5.1

 

 

 

Output setpoints (PID)5.2

 

 

 

Output setpoints

5.3

 

 

 

Simulated tests

5.4

 

 

 

 

 

 

 

 

 

 

Display

Outputs

Alarms

Exit

More Enter

PRESS

ARROWS

b

Diagnostics

Security

Exit

More Enter

CONFIGURE MENU SEC- TION

Display 5.5

Outputs 5.6

Alarms 5.7

Diagnostics 5.8

Security 7.1

FIGURE A-1. Outline of Menu Levels for ORP

74

Page 80
Image 80
Emerson 54e pH/ORP instruction manual Output setpoints, Simulated tests