MODEL 54e pH/ORP

SECTION 5.0

 

SOFTWARE CONFIGURATION

TABLE 5-1. pH Settings List (continued)

ITEM

RANGES

FACTORY SETTINGS

USER SETTINGS

2b. Output 1 Setup (PID)

 

 

 

 

(a) Setpoint

-2 to 16 pH or -15 to 130°C

 

7 pH

_______

(b) Proportional

0-299.9%

 

100.0%

_______

(c) Integral

0-2999 sec

 

0 sec

_______

(d) Derivative

0-299.9%

 

0.0%

_______

(e) LRV (4 mA)

-2 to 16 pH or -15 to 130°C

 

2 pH

_______

(f) URV (20 mA)

-2 to 16 pH or -15 to 130°C

 

0 pH

_______

3. Output 2 Control

 

 

 

 

(a) Output 2 Measurement

Process/Temp/Glass Imp/Ref Imp

Temperature

_______

(b) Output 2 Control Mode

Normal/PID

 

Normal

_______

4a. Output 2 Setup (Normal)

 

 

 

 

(a) Current Range

4-20 mA/0-20 mA

 

4-20 mA

_______

(b) Dampening

0-255 Sec

 

0 Sec

_______

(c) Hold Mode

Last value/Fixed value

 

Last value

_______

(d) Fixed Hold Value (if (c) Fixed)

0-22 mA

 

21 mA

_______

(e) Fault value

0-22 mA

 

22 mA

_______

4b. Output 2 Setup (PID)

 

 

 

 

(a) Setpoint

-2 to 16 pH or -15 to 130°C

 

7 pH

_______

(b) Proportional

0-299.9%

 

100.0%

_______

(c) Integral

0-2999 sec

 

0 sec

_______

(d) Derivative

0-299.9%

 

0.0%

_______

(e) LRV (4 mA)

-2 to 16 pH or -15 to 130°C

 

2 pH

_______

(f) URV (20 mA)

-2 to 16 pH or -15 to 130°C

 

0 pH

_______

5. Hold (Outputs and Relays)

Disable/Enable/ 20 min timeout

Disable feature

_______

C. Alarms (Section 5.7)

 

 

 

 

1. Alarm 1 Control

 

 

 

 

(a) Activation Method

Process/Temp

 

Process

_______

(b) Alarm 1 Control Mode

Normal/TPC

 

Normal

_______

2a. Alarm 1 Setup (Normal)

 

 

 

 

(a) Configuration

Low alarm/High alarm/Off

 

Low

_______

(b) Hysteresis (deadband)

0 - 5.00 pH

 

0.01 pH

_______

(c) Delay Time

0-99 sec

 

0 sec

_______

(d) Relay Fault

Open/Closed/None

 

None

_______

2b. Alarm 1 Setup (TPC)

 

 

 

 

(a) Setpoint

-2 to 16 pH or -15 to 130°C

 

7 pH

_______

(b) Proportional

0-299.9%

 

100.0%

_______

(c) Integral

0-2999 sec

 

0 sec

_______

(d) Derivative

0-299.9%

 

0.0%

_______

(e) Time Period

10-2999 sec

 

30 sec

_______

(f) LRV (100% On)

-2 to 16 pH or -15 to 130°C

 

2 pH

_______

(g) URV (100% Off)

-2 to 16 pH or -15 to 130°C

 

0 pH

_______

(h) Relay Fault

None/Open/Closed

 

None

_______

3. Alarm 2 Control

 

 

 

 

(a) Activation Method

Process/Temp

 

Process

_______

(b) Alarm 2 Control Mode

Normal/TPC

 

Normal

_______

4a. Alarm 2 Setup (Normal)

 

 

 

 

(a) Configuration

Low alarm/High alarm/Off

 

High

_______

(b) Hysteresis (deadband)

0 - 5.00 pH

 

0.01 pH

_______

(c) Delay Time

0-99 sec

 

0 sec

_______

(d) Relay Fault

Open/Closed/None

 

None

_______

4b. Alarm 2 Setup (TPC)

 

 

 

 

(a) Setpoint

-2 to 16 pH or -15 to 130°C

 

7 pH

_______

(b) Proportional

0-299.9%

 

100.0%

_______

(c) Integral

0-2999 sec

 

0 sec

_______

(d) Derivative

0-299.9%

 

0.0%

_______

(e) Time Period

10-2999 sec

 

30 sec

_______

(f) LRV (100% On)

-2 to 16 pH or -15 to 130°C

 

2 pH

_______

(g) URV (100% Off)

-2 to 16 pH or -15 to 130°C

 

0 pH

_______

(h) Relay Fault

None/Open/Closed

 

None

_______

 

 

 

 

 

Continued on the following page

23

Page 29
Image 29
Emerson 54e pH/ORP instruction manual Ranges Factory Settings User Settings