MODEL 54e pH/ORP

SECTION 5.0

 

SOFTWARE CONFIGURATION

5.6 CHANGING OUTPUT PARAMETERS (continued)

Fixed Hold: 21.00mA

Fault: 22.00mA

"Proportional" is short for Proportional Band and indicates the range over which control is being used. It is the opposite of the process gain. Smaller values provide tighter control.

"Integral" is the number of seconds over which deviations from the set- point are integrated to remove continuing offsets. Smaller values provide higher response.

"Derivative" is a form of control that resists all changes in readings. Higher readings increase the derivative function. Use caution in setting the derivative value to prevent process oscillation.

More information regarding PID control can be found in Section 7.0.

ExitEdit Setting these parameters may require some trial and error and should be tested while the process is being supervised to prevent future upsets.

The rest of the PID output setup parameters are identical to those used for normal outputs. See step 4a for details .

Hold Feature Setup

Output 2 control

Output 2 setup

Hold feature setup

ExitEnter

Hold: Disable feature

ExitEdit

5.The Hold feature is used to prevent problems that may occur during calibration if the current outputs are used for control. The feature is turned on (enabled) here and is specifically configured in step 4. The controller starts out with the hold feature turned off (disabled).

To enable the Hold feature, obtain the screen to the left with the hold feature setup highlighted (see steps 1 and 2 for exact instructions). Press Enter 4 and the screen below will appear.

Press the Edit 4 key to enable changes. Options include Disable, Enable, and 20 minute timeout. When 20 minute timeout is selected, the hold mode will automatically disengage after being on for 20 minutes. Selecting Enable or 20 minute timeout does not actually put the controller in hold, but rather allows putting the controller in hold when calibration is conducted.

7.00

pH

When the hold feature has been enabled, this Hold Mode Screen will

 

 

 

 

 

appear when the Calibrate routine is entered. Possible actions are Exit

 

 

1 which cancels the calibration, Enter 4 which enters the calibrate

Hold mode: Off

 

menu without putting the controller in hold, and Edit 4 which allows

 

 

turning Hold Mode On. Note that when hold has been enabled, this

Exit

Enter

screen requires pushing Cont 3 to enter and leave the calibrate menu.

35

Page 41
Image 41
Emerson 54e pH/ORP Fixed Hold 21.00mA, Fault 22.00mA, Output 2 control Output 2 setup, Hold feature setup