MODEL 54e pH/ORP | APPENDIX A |
| ORP CONFIGURATION |
APPENDIX A
ORP CONFIGURATION
QUICK STARTUP (ORP) | CONFIGURING THE CONTROLLER |
The Model 54e pH/ORP controller comes
To switch to an ORP controller and perform a quick startup:
1.Connect the ORP sensor to the controller. Wire to controller per Figure
2.Apply primary power (Figure
3.Change the controller to ORP operation. Refer to Section 5.3, or:
A.Press any key to display the main menu.
B.Move cursor until Program is highlighted.
C.Press Enter (F4).
D.Highlight "Configure" and press Enter (F4).
E.Highlight "Display" and press Edit (F4).
F.Use the arrow key to change "Measure" from pH to ORP and press Save (F4). Warning screen appears - press Cont (F3).
4.Complete the ORP standardize instructions).
5.Preset alarm and output settings are listed below: If these values are unacceptable, refer to Section A.1 for a detailed configuration proce- dure. For a summary of the complete factory set- tings see Table
Alarm 1 | (low alarm): | ||
Alarm 2 | (high alarm): | +1400mV | |
Alarm 3 | (high alarm): | +1400mV | |
Output 1 | (4mA setpoint): | ||
Output 1 | (20mA setpoint): | +1400mV |
Configuration of the ORP settings can be done at any time after the controller has been set to ORP mode (step 3 in Quick Startup on the left). Procedures for these operations follow the ORP Standardization instructions in this appendix.
Before attempting any changes refer to the configu- ration setup list shown in Table
On initial configuration, it is recommended that the settings be entered in the order shown on the work- sheet. This will reduce the chance of accidentally omitting a needed setting.