Rosemount 3300 Series

'HDG￿=RQHV

The measuring range depends on probe type and properties of the product.

 

The 8SSHU￿'HDG￿=RQH is the minimum measurement distance between the

 

upper reference point and the product surface. The Upper Dead Zone varies

 

between 4 - 20 in. (0.1 and 0.5 m) depending on probe type and product.

At the end of the probe the measuring range is reduced by the /RZHU￿'HDG￿ =RQH. The /RZHU￿'HDG￿=RQH￿also varies depending on probe type and

product.

Figure 2-5 illustrates how the measuring range is related to the Dead Zones:

Figure 2-5. Dead Zones

Upper Reference Point

8SSHU￿'HDG￿=RQH

20mA

%

 

100-

0D[LPXP￿

Range 0

0HDVXULQJ￿5DQJH

 

ZONES

/RZHU 'HDG￿=RQH

 

4mA

DEAD

 

Table 2-2￿￿;￿. Dead ￿￿￿(￿￿￿￿!￿￿￿￿￿￿￿;￿￿(￿￿￿￿Zones for different probe types

 

 

 

 

 

 

 

Upper Dead

2

3.9 in. (10 cm)

7.9 in. (20 cm)

15.7 in. (40 cm)

13.8 in. (35 cm)

19.7 in. (50 cm)

Zone

 

 

 

 

 

 

80

3.9 in. (10 cm)

7.9 in. (20 cm)

11.8 in. (30 cm)

7.9 in. (20 cm)

11.8 in. (30 cm)

 

 

 

 

 

 

 

Lower Dead

2

2 in. (5 cm)

2.8 in. (7 cm)

5.9 in. (15 cm)

3.9 in. (10 cm)

4.7 in. (12 cm)

Zone

 

 

 

 

 

 

80

1.2 in. (3 cm)

2 in. (5 cm)

2 in. (5 cm)

2 in. (5 cm)

2 in. (5 cm)

 

 

 

 

 

 

 

127(

The measurement accuracy is reduced in the Dead Zones. It may even be impossible to make any measurements at all in those regions. Therefore the 4-20 mA set points should be configured outside the Dead Zones.