Emerson Process Management rosemount 3300 series 966 +$5$&75,67,&6, 7DQN6KDSH, Pxovlrq/D\Huv

Models: rosemount 3300 series

1 128
Download 128 pages 23.28 Kb
Page 20
Image 20

Rosemount 3300 Series

Figure 2-7. Reduction of maximum measuring range for

Flexible Twin Lead probe

_SCALE

 

 

INTERFACE_REDUCTION

￿￿￿￿￿(￿￿￿;￿￿￿￿￿<￿￿￿￿￿*￿!￿￿￿￿￿￿￿￿￿;￿￿7￿￿5￿8*￿￿$￿ ￿￿￿￿￿￿￿￿￿￿￿￿￿￿￿5HIHUHQFH￿ ￿0DQXDO

00809-0100-4811, Rev AA

October 2002

+0￿1

*￿￿￿￿￿￿￿<￿￿￿￿￿￿￿￿￿￿￿￿(￿￿￿￿￿￿￿￿￿7￿￿5￿8

(PXOVLRQ￿/D\HUV

Sometimes there is an emulsion layer (mix of the products) between the two products which, depending on its characteristics, will affect interface measurements.

For guidelines on emulsion situations, consult factory.

9(66(/￿

&+$5$&7(5,67,&6

+HDWLQJ￿&RLOV￿￿$JLWDWRUV

7DQN￿6KDSH

The Rosemount 3300 Series Radar Transmitter is relatively insensitive to objects in the tank since the radar signal is transmitted along a probe.

Avoid physical contact between probes and agitators as well as applications with strong fluid movement unless the probe is anchored. If the probe can move within 1 ft (30 cm) away from any object, such as an agitator, during operation then probe tie-down is recommended.

In order to stabilize the probe for side forces, it is possible to hang a weight at the probe end (flexible probes only) or fix/guide the probe to the tank bottom.

The guided wave radar transmitter is insensitive to the tank shape. Since the radar signal travels along a probe, the shape of the tank bottom has virtually no effect on the measurement performance. The transmitter handles flat or dish-bottom tanks equally well.

2-10