63(&,),&$7,216

 

 

 

 

 

 

 

 

 

 

 

 

 

Product

 

Rosemount Series 3300 Guided Wave Radar Level and Interface

 

 

 

Transmitter;

 

 

 

Model 3301 for Level (Interface available for fully immersed probe).

 

 

 

Model 3302 for Level and Interface.

 

 

 

 

 

Measurement principle

 

Time Domain Reflectometry (TDR).

 

Reference Conditions

 

Twin Lead probe, 77 °F (25 °C) water

 

 

 

 

 

Microwave Output Power

 

Nominal 50 W, Max. 2 mW.

 

CE-mark

 

Complies with applicable directives (R&TTE, EMC, ATEX)

 

 

 

 

 

 

 

 

 

 

 

 

 

Integral Display

 

The integral display toggles between the following variables: level,

 

 

 

distance, volume, internal temperature, interface distance, interface level,

 

 

 

peak amplitudes, interface thickness, percent of range, analog current

 

 

 

output.

 

 

 

Note! The Integral Display can not be used for configuration purposes.

 

 

 

 

 

Output Units

 

For Level, Interface and Distance: ft, inch, m, cm or mm.

 

 

 

For Volume: ft3, inch3, US gals, Imp gals, barrels, yd3, m3 or liters.

 

Output Variables

 

Model 3301: Level, Distance (to product surface) and Volume. With fully

 

 

 

immersed probe: Interface Level and Interface Distance.

 

 

 

Model 3302: Level, Distance (to product surface), Volume, Interface Level,

 

 

 

Interface Distance and Upper Product Thickness.

 

 

 

 

 

HART device for remote confifguration

 

Rosemount hand-held communicator Model 275.

 

PC for remote configuration

 

Radar Configuration Tools software package.

 

 

 

Rosemount AMS software.

 

 

 

 

 

 

 

 

 

 

 

 

 

Power supply

 

Loop-powered (2-wire), 11 - 42 VDC (11 -30 VDC in IS applications, 16-42

 

 

 

VDC in Explosion Proof/Flame Proof applications).

 

 

 

 

 

Output

 

Analog 4 - 20 mA, HART.

 

Signal on alarm

 

Standard: Low=3.75 mA. High=21.75 mA.

 

 

 

Namur NE 43: Low=3.60 mA. High=22.50 mA.

 

 

 

 

 

IS parameters

 

Ui = 30 V,li= 130 mA, Pi = 1 W, Li=0, Ci=0.

 

Cable entry

 

½ - 14 NPT for cable glands or conduit entries.

 

 

 

Optional: M20 x 1.5 conduit adapter or PG 13.5 conduit adapter.

 

 

 

 

 

Output Cabling

 

Twisted shielded pairs, 18-12 AWG..

 

 

 

 

www.rosemount.com