Emerson Process Management rosemount 3300 series Example 1 amplitude threshold T2 is too high

Models: rosemount 3300 series

1 128
Download 128 pages 23.28 Kb
Page 83
Image 83

5HIHUHQFH￿0DQXDO￿

00809-0100-4811, Rev AA October 2002

$PSOLWXGH￿7KUHVKROG￿

6HWWLQJV

Figure 6-10. Example 1: amplitude threshold T2 is too high.

Figure 6-11. Example 2: amplitude threshold T2 is too low.

Rosemount 3300 Series

The amplitude thresholds are automatically adjusted to appropriate values in order to filter out noise and other non-valid measurements from the measurement signal.

The amplitude of the measurement signal, i.e. the amplitude of the signal that is reflected by the product surface, is related to the actual dielectric constant of the product. The amplitude threshold that is used by the transmitter is

based on the parameter configuration of the current product dielectric constant (see 6HFWLRQ ￿￿￿%DVLF￿&RQILJXUDWLRQ). Normally no other threshold

adjustment is needed, but if the transmitter still does not track the product surface correctly it may be necessary to adjust the threshold values.

The Radar Configuration Tool (RCT) has a plot function allowing you to view the reflections along the probe.

If the amplitude threshold is too high the product level is not detected as illustrated in Figure 6-10.

 

100

 

(+￿￿!￿"￿￿￿;￿"￿￿￿￿￿￿

 

 

250

 

 

80

 

 

 

 

T3

200

HIGH

 

60

 

 

 

 

 

 

 

 

 

 

 

 

Amplitude

-20

 

 

 

T2

 

 

THRESHOLD

 

40

 

 

 

 

 

150

 

 

20

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

0

 

 

 

 

 

100

 

 

 

 

 

 

 

 

 

 

 

T1

 

 

 

 

50

WAVEFORMPLOT

 

-40

 

 

 

 

 

 

 

 

 

 

 

 

 

 

P1

 

 

 

 

 

 

-60

 

 

 

 

 

 

 

 

 

 

 

 

0

 

 

 

 

 

 

 

 

 

 

0

100

200

300

400

500

600

 

 

 

 

 

Distance (samples)

 

 

 

If there are disturbing objects in the tank the threshold must be carefully set in order to avoid locking on the wrong amplitude peak. In Figure 6-11the transmitter has locked on a peak above the actual product surface, i.e. a disturbance was interpreted as the product surface, whereas the actual product surface was interpreted as an interface or the probe end.

 

100

$￿￿￿￿￿￿￿￿￿;￿￿￿￿￿￿￿￿￿

 

P3

￿￿￿￿￿￿￿￿￿￿￿￿￿￿

T3

 

250

LOW

 

80

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

60

 

 

 

 

 

 

 

200

THRESHOLDWAVEFORMPLOT

 

 

 

 

 

 

 

 

 

Amplitude

40

 

 

 

P2

 

 

 

150

20

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

T2

 

 

 

 

 

 

 

 

 

 

 

 

 

 

0

 

 

 

 

 

 

 

100

 

 

 

 

 

 

 

 

 

 

 

 

-20

 

T1

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

50

 

 

-40

 

 

 

 

 

 

 

 

 

 

P1

 

 

 

 

 

 

 

 

-60

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

0

 

 

 

0

100

200

300

400

500

600

 

 

 

 

 

 

 

 

 

 

Distance (samples)

 

 

 

 

6-13

Page 83
Image 83
Emerson Process Management rosemount 3300 series manual Example 1 amplitude threshold T2 is too high