Rosemount 3300 Series

5HIHUHQFH￿0DQXDO

00809-0100-4811, Rev AA October 2002

Figure 6-12. Waveform plot after threshold T2 was adjusted

By adjusting the amplitude threshold T2 the product surface is detected correctly as illustrated in Figure 6-12.

 

100

 

 

P2

￿￿￿￿￿￿￿￿￿￿￿￿￿￿￿￿￿￿￿￿(+￿￿￿￿￿￿￿￿￿￿￿￿￿￿￿￿￿￿￿250

ADJUSTED

 

 

 

 

 

80

T3

 

 

 

 

 

 

 

 

 

 

 

 

 

200

 

 

60

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Amplitude

40

T1

 

 

 

 

T2

150

THRESHOLD

 

 

 

 

 

 

 

 

 

 

 

 

 

 

20

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

0

 

 

 

 

 

 

100

 

 

 

 

 

 

 

 

 

 

 

-20

 

 

 

 

 

 

 

 

 

-40

 

 

 

 

 

 

50

WAVEFORMPLOT

 

 

P1

 

 

 

 

 

 

 

 

 

 

 

 

 

 

-60

 

 

 

 

 

 

 

 

 

 

 

 

 

 

0

 

 

 

 

 

 

 

 

 

 

 

0

100

200

300

400

500

 

600

 

 

 

 

 

Distance (samples)

 

 

 

 

To adjust the amplitude thresholds select HART command [1, 4, 6, 2]

or

1.Start the Radar Configuration Tool (RCT).

2.Choose the 'HYLFH￿&RPPDQGV option from the View menu.

3.Open the 'HWDLOV folder.

4.Click the 6HW￿1RPLQDO￿7KUHVKROGV￿option.

The thresholds T2 and T3 should be set to about 50% of the measured signal amplitude for the product surface and the interface peaks, respectively.

127(

Amplitude thresholds should not be set to values less than 3.

127(

Check that the dielectric constant parameter setting is reasonably close to the actual dielectric constant value of the upper product before changing the amplitude thresholds.

127(

Default Amplitude thresholds can be set by typing 0 as the new threshold value.

6-14

Page 84
Image 84