2-7
Trimmingthe analog output is a calibration of the output c ircuitry,by setting
the4 and 20 mA points. Once the 4 and 20 mA points are set, all
intermediatevalues are automatically adjusted. See Menu Trees in Section
3fora corresponding menu tree.
Fromthe Online menu, select Device setup.
Fromthe Device setup menu, selec t Diag/
Service.
Fromthe Diag/Service m enu, select
Calibration.
Fromthe Calibration menu, selectTrim
analogoutput.
Fromthe Trim Analog Output menu, select
thedesired option and follow the on-line
instructions.
12
50 0