Chapter 6. Programming Meter Settings
124 Transport® Model PT878 Portable Liquid Flowmeter User’s Manual
6.9 Changing Communications Settings (cont.)
1. From the Meter menu, scroll to the Communications entry and press
[ENTER]. The Communications window appears similar to Figure 83
below.
Figure 83: Communications Window
2. The first prompt asks for the node identification number, which can be
any number from 1 to 240. Do NOT change the node ID unless
instructed by GE.
a. Press [ENTER] to open the text box.
b. Use the numeric keys to enter the node ID.
c. Press [ENTER] to confirm the entry.
3. The next prompt asks you to select the communications interface from
two options: IrDA (Infrared Data Association) and the IR232 virtual
interface. You must use IrDA when using an IR printer or transfer
application such as QuickBeam. Use the [W] and [X] keys to scroll to
the desired selection, and press [ENTER].
Communications
Node ID 2
Comm Interface IrDA IR 232
Baud Rate 9600 bps
Parity None
Stop Bits
Data Bits
12
78
Cancel OK