Chapter 3. Programming Site Data
Transport® Model PT878 Portable Liquid Flowmeter User’s Manual 65
3.11.1 Entering Correction Factors
The final option in the Program Menu, Correction Factors, allows you to
enter and modify three correction factors: Reynolds Correction, Kinematic
Viscosity and Calibration Factor. To enter the Correction Factors option,
scroll to the Correction Factors entry on the Program Menu and press
[ENTER]. The screen appears similar to Figure 35 below. To step through
each parameter, press the [T] key. Refer to Figure152 on page 236 of
Appendix A, Menu Maps.
Figure 35: Reynolds Correction Window
3.11.2 Entering Reynolds Correction
The default for Reynolds Correction is “On.” This correction factor should
be on in most applications, including all those that utilize clamp-on
transducers. It makes a small adjustment to the flow rate reported by the
PT878, based on the Kinematic Viscosity. Reynolds Correction is necessary,
as the velocity of the fluid measured along a diametrical path must be
related to the total area average velocity over the entire pipe cross-section.
Reynolds Correctio Inputs
Correction Factors
Cancel OK
Calibration Factor
Re Correction On Off
K/V (v) m3/s
Data Source
K/V Factor Single Table
Edit Table
10