Chapter 3. Programming Site Data
68 Transport® Model PT878 Portable Liquid Flowmeter User’s Manual
3.11.3 Entering a Calibration Fact or (cont.)
3. The following steps depend on whether you select a single value or a
table.
If you entered Single:
The prompt asks for a value. This feature enables a single multiplier to be
applied to the flow rate reported by the PT878.
a. Press [ENTER] to open the text box.
b. Use the numeric keys to enter the desired value. Generally, if you
have enabled the Reynolds Correction factor, the correction factor
should be set to 1.00. Otherwise, the typical factor is between 0.5
and 2.00.
c. Press [ENTER] to confirm your selection.
You have completed entering data for a single Calibration Factor.
To confirm the entries and return to Operate mode, press [F3] (OK). The
PT878 returns to Operate Mode.
To leave the window without confirming the entries, press [F2] (Cancel)
or the [ESC] key. The PT878 returns to Operate Mode.