GE PT878 user manual If you selected Pulse Totalizer

Models: PT878

1 297
Download 297 pages 27.94 Kb
Page 74
Image 74

Chapter 3. Programming Site Data

3.10Entering the Digital Output (cont.)

a.Press [ENTER] to open the drop-down menu.

b.Scroll to the desired function.

c.Press [ENTER] to confirm your selection.

Note: If you select Off, you will not be able to access any other parameters in this option. If you select Test Points or Gate Input, the program goes at once to Step 4.

2.The next prompt asks you to select the data source for the digital output. The choices vary with the function. For Pulse Totalizer (shown in Figure 31 on page 57, the available choices include:

Forward Totalizer

Reverse Totalizer

Forward Energy

Reverse Energy

a.Press [ENTER] to open the data source selection window, shown in Figure 30 on page 55.

b.Scroll to the desired source.

c.Press [SEL] to confirm your selection.

d.The prompt then moves to a list of unit types. (The available units depend on the selection made at the Data Source prompt.) Scroll to the desired output unit.

e.Press [F3] (OK) to confirm your selection.

3.The last set of parameters that appears depends on the selection you made at the Function prompt.

If you selected Pulse Totalizer:

The prompt asks for the units/pulse, the pulse width (in microseconds), and the polarity. (Figure 31 on page 57 illustrates a Digital Output window configured for the Pulse Totalizer function.) For the Units/Pulse and Pulse Width parameters:

58

Transport® Model PT878 Portable Liquid Flowmeter User’s Manual

Page 74
Image 74
GE PT878 user manual If you selected Pulse Totalizer