Chapter 3. Programming Site Data
Transport® Model PT878 Portable Liquid Flowmeter User’s Manual 37
3.5 Entering Fluid Types and Speeds
To access the Fluid option, scroll to the Fluid entry on the Program Menu
and press [ENTER]. (If you are already in the Transducer/Pipe form, press
the [X] arrow key to reach the Fluid window, and press [ENTER].)The
screen appears similar to Figure 19 below. To step through each parameter,
press the [T] key. Refer to Figure145 on page 229 of Appendix A, Menu
Maps.
Figure 19: Fluid Type Window
1. The first prompt asks you to select whether or not you want Tracking
Windows. These windows are used to detect the receive signal when
you are unsure of the fluid soundspeed. (Default operation is “No.”)
a. Use the [W] and [X] keys to scroll to the appropriate radio button.
b. Press [ENTER] to confirm your selection.
Fluid
Cancel OK
Transducer/Pipe
Transducer Pipe Path
Fluid
Lining
Tracking Windows? No Yes
Fluid
Sound Speed
Speed Min
Speed Max
Glycol
Water (2C)
1496
1350
1650
10
m/s
m/s
m/s
%
Tem p °F
77