Chapter 3. Programming Site Data

3.3Entering Pipe Parameters (cont.)

d.If you have selected “Other,” the meter prompts you to enter the soundspeed. Use the numeric keys to type the desired soundspeed in the text box, and press [ENTER] to confirm the choice.

Note: If the “Other” Pipe soundspeed entered is too large, given the previously entered Wedge soundspeed and angles, an error message similar to Figure 16 below will appear. Press [F3] (OK) (the error message disappears), and enter another soundspeed within the range specified.

Transducer/Pipe

Invalid Value - Too High

100000 is Too High

The valid range is from: 3330.052 to 24000.000 ft/s

Cancel

OK

Figure 16: Soundspeed Error Message Window

Note: When the corrected soundspeed is entered, a message similar to Figure 17 on page 33 appears. Press [F3] (OK).to return to the Site Menu.

32

Transport® Model PT878 Portable Liquid Flowmeter User’s Manual

Page 48
Image 48
GE PT878 user manual Transducer/Pipe Invalid Value Too High, Is Too High Valid range is from 3330.052 to 24000.000 ft/s

PT878 specifications

The GE PT878 is a state-of-the-art portable ultrasonic flow meter designed for the precise measurement of liquid flow in various applications. As part of General Electric’s renowned instrumentation lineup, the PT878 combines innovative technology with user-friendly features, making it an invaluable tool for engineers, technicians, and plant operators.

One of the standout characteristics of the PT878 is its non-invasive measurement capability. The meter utilizes clamp-on transducer technology, allowing it to be installed on the exterior of the pipe, eliminating the need for invasive procedures which can disrupt the system or require pipe cutting. This non-contact approach not only simplifies installation but also enhances the device’s versatility, as it can be applied to a wide range of pipe sizes and materials.

The PT878 leverages advanced digital signal processing and sophisticated algorithms to provide accurate and reliable flow measurements. It operates using the time-of-flight method, whereby ultrasonic signals are sent and received through the fluid in the pipeline to determine the flow rate. This technology allows for remarkable precision, ensuring that operational efficiency can be maintained.

Moreover, the PT878 is equipped with features such as a bright, easy-to-read display, intuitive navigation, and robustness against harsh environmental conditions. Its compact and lightweight design enhances portability, making it suitable for field applications. The meter operates on battery power, providing a long-lasting solution for remote monitoring and temporary installations.

Another impressive aspect of the PT878 is its ability to measure not only water but also a variety of liquids, from hydrocarbons to chemicals. It supports multiple measurement modes, including volumetric flow and velocity, offering a level of flexibility that caters to diverse industrial needs.

In addition to flow measurement, the PT878 offers data logging capabilities, allowing users to store and analyze flow data over time. This function is vital for understanding trends, diagnosing issues, and optimizing system performance. Users can easily transfer data to computers for comprehensive analysis, enhancing operational insights.

With its combination of innovative technology, user-friendliness, and robust measurements, the GE PT878 ultrasonic flow meter stands out as a comprehensive solution for industries requiring reliable liquid flow measurement. Whether in a manufacturing plant, a water treatment facility, or any application requiring fluid monitoring, the PT878 delivers exceptional performance and accuracy.