Chapter 3. Programming Site Data
54 Transport® Model PT878 Portable Liquid Flowmeter User’s Manual
3.9 Entering the Analog Output
The Analog Output option enables you to enter information to set up output
parameters. To enter the option, scroll to the Analog Output entry on the
Program Menu and press [ENTER]. The screen appears similar to Figure 29
below. To step through each parameter, press the [T] key. Refer to
Figure 149 on page233 of Appendix A, Menu Maps.
Figure 29: Analog Output Window
1. The first prompt enables you to select a range to send a current signal to
a recording device.
a. Press [ENTER] to open the drop-down menu.
b. Scroll to the desired output from three choices: Off, 0-20 mA, and
4-20 mA.
c. Press [ENTER] to confirm your selection.
Note: If you select “Off,” you will not be able to access any other
parameters in this option.
Analog Output
Cancel OK
Function
Data Source
Zero
On Error
0-20 mA
Veloci ty
Span
Hold Last Value
0
Meters/sec
25.5
m/sm/s