Chapter 3. Programming Site Data

3.9Entering the Analog Output (cont.)

3.The next prompt asks you to enter the zero (base) value for the analog output. This value represents the 0/4 mA output (in flow units).

a.Press [ENTER] to open the text box.

b.Use the numeric keys to enter the desired value.

c.Press [ENTER] to confirm your selection.

4.The next prompt asks you to enter the span (full scale) value for the analog output. This value represents the 20 mA output (in flow units).

a.Press [ENTER] to open the text box.

b.Use the numeric keys to enter the desired value.

c.Press [ENTER] to confirm your selection.

5.The last prompt, On Error, asks you to select how the PT878 will handle the analog outputs in the event of a fault condition. The meter offers three alternatives:

Hold Last Value (hold the last good reading)

Force Low (force the reading to 0 or 4 mA)

Force High (force the reading to 20 mA).

a.Press [ENTER] to open the drop-down menu.

b.Scroll to the desired selection.

c.Press [ENTER] to confirm your selection.

You have completed entering data in the Analog Output option.

To confirm the entries and return to Operate mode, press [F3] (OK). The PT878 returns to Operate Mode.

To leave the window without confirming the entries, press [F2] (Cancel) or the [ESC] key. The PT878 returns to Operate Mode.