Menu Reference

Front Panel Menu Reference "System/Inputs" Menu

System

Inputs

The softkey menu structured to this hardkey affects the system setup and the channel setup.

Press this hardkey to access the “System/Inputs” menu. This menu allows you to set up a variety of conditions for your power meter, for example: HP-IB address, sensor calibration tables, averaging, duty cycle, range, offset, service. It also allows you to review errors.

HP-IB

Press this softkey to access a menu which allows you to set the HP-IB address and select the HP-IB command set you want to use.

HP-IB Addr

Press this softkey to set the HP-IB address of your power meter between 0 and 30. The default value is 13. A pop up

window appears on the screen. Use the , , and hardkeys to change the value. To confirm your choice

press Enter . Refer to “HP-IB Address”, on page 2-42 for further information.

Sensor Cal Tables

Press this softkey to access a menu and screen which allows you to

select and edit sensor calibration tables. Use the and hardkeys to scroll through the displayed sensor calibration tables. Sensor calibration tables are only required for the HP 8480 series power sensors.

Edit Cal Table

Press this softkey to access a menu and screen which allows you to modify, add and remove frequency and calibration factors from the currently selected table and modify the table

name. Use the , , and hardkeys to move between the name, frequency and calibration factors. Press Change, Insert or Delete as follows:

Change

Press this softkey to modify the highlighted parameter, which is either a frequency, calibration factor or table name.

A pop up window displays the parameter. Use the ,

, and hardkeys to change the parameter value. To confirm your choice press Enter .

3-18

HP EPM-442A User’s Guide