Power Meter Operation

Calibrating the Power Meter

Table 2-1: Connecting the HP 8480 Series Power Sensors During Calibration

Sensor

Connection Requirements

Model

 

 

 

 

 

HP 8481A

These power sensors connect directly to the reference calibrator.

HP 8481H

 

HP 8482A

 

HP 8482H

 

 

 

HP 8481D

Prior to the power meter being calibrated an HP 11708A 30 dB

HP 8484A

reference attenuator should be connected between the power

 

sensor and the reference calibrator. This attenuator must be

 

removed from the power sensor input prior to making

 

measurements.

 

 

HP 8483A

This power sensor requires a 75 Ω (f) to 50 Ω (m) N-Type adapter

 

(1250-0597) to connect to the reference calibrator. This adapter

 

must be removed from the power sensor input prior to making

 

measurements.

 

 

HP R8486A

The waveguide power sensors have two connectors. The N-Type

HP Q8486A

connector is the one which is used to calibrate the power meter.

HP R8486D

 

HP Q8486D

 

HP W8486D

 

 

 

HP 8481B

These power senors are configured with an attenuator. Prior to

HP 8482B

the power meter being calibrated, this attenuator must be

 

removed. The attenuator must be reconnected prior to making

 

measurements.

 

 

HP 8485A

This power sensor requires an APC 3.5 (f) to 50 Ω (m) N-Type

 

adapter (08485-60005) to connect to the reference calibrator.

 

 

HP 8485D

Prior to the power meter being zeroed and calibrated an

 

HP 11708A 30 dB reference attenuator and an APC 3.5 (f) to

 

50 Ω (m) N-Type adapter (08485-60005) should be connected

 

between the power sensor and the reference calibrator. This

 

attenuator must be removed from the power sensor input prior to

 

making measurements.

 

 

HP 8487A

This power sensor requires an APC 2.4 (f) to 50 Ω (m) N-Type

 

adapter (08487-60001) to connect to the power meter.

 

 

HP 8487D

Prior to the power meter being zeroed and calibrated an

 

HP 11708A 30 dB reference attenuator and an APC 2.4 (f) to

 

50 Ω (m) N-Type adapter (08487-60001) should be connected

 

between the power sensor and the reference calibrator. This

 

attenuator must be removed from the power sensor input prior to

 

making measurements.

 

 

HP EPM-442A User’s Guide