Frequency Cal Fac

Power Meter Operation

Making Measurements with the HP 8480 Series Power Sensors

Press A Ref CF . Use the , , and hardkeys to enter 99.8. Press % .

Connect the power sensor to the POWER REF output.

Press Cal A .

• Press , A Cal Fac. Use the , , and hardkeys to enter 97.8. Press % .

Connect the power sensor to the signal to be measured.

The measurement result is now displayed.