Power Meter Operation

Making Measurements using Sensor Calibration Tables

The power meter is shipped with a set of predefined sensor calibration tables. The data in these tables is based on statistical averages for a range of Hewlett-Packard power sensors.

These power sensors are:

DEFAULT1

HP 8481A

HP 8482A2

HP 8483A

HP 8481D

HP 8485A

R8486A

Q8486A

R8486D

HP 8487A

There are also ten sensor calibration tables named CUSTOM_0 through CUSTOM_9 which do not contain any data when the power meter is shipped from the factory.

You cannot delete any of the 20 sensor calibration tables. Each sensor calibration table can contain a maximum of 80 frequency points each. If you want a new sensor calibration table you must edit an existing one.

To edit a table:

1.Select one using the and hardkeys to scroll through the displayed tables. When the table you want to edit is highlighted press Edit Cal Table . The “Edit Cal” screen is displayed as shown in Figure 2-2 on page 2-16. In this screen you can modify,

add and remove frequency and calibration factors and also edit the

table name. Use the , , and hardkeys to move between the table name and its frequency and calibration factors.

A frequency in the range of 0.1 MHz to 999.999 GHz can be entered.

1.DEFAULT is a sensor calibration table in which the reference calibration factor and calibration factors are 100%. This sensor calibration table can be used during the performance testing of the power meter.

2.The HP 8482B and HP 8482H power sensors use the same data as the

HP 8482A.

HP EPM-442A User’s Guide

2-17