Power Meter Operation
Making Measurements using Sensor Calibration Tables
“State” field indicates if any sensor calibration tables are currently selected. The “Sensor Tbls” screen is displayed as shown in Figure
Figure
| Making the Measurement | ||||||||||||||
| To make the power measurement, you zero and calibrate the power meter, | ||||||||||||||
| then set it for the frequency of the signal you want to measure. The power | ||||||||||||||
| meter automatically selects the calibration factor from the sensor | ||||||||||||||
| calibration table. | ||||||||||||||
|
|
|
|
|
|
|
|
|
|
|
|
|
|
| |
Note | The following procedure details the calibration for channel A. To | ||||||||||||||
| calibrate channel B use the same procedure using the equivalent | ||||||||||||||
| channel B softkeys. To calibrate both channels, zero the power | ||||||||||||||
| meter using |
|
|
|
|
|
|
| in step “2”, then follow this procedure for | ||||||
| Zero Both | ||||||||||||||
| channel A then repeat step “3” through to step “9” using the | ||||||||||||||
| equivalent channel B softkeys. | ||||||||||||||
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
| 1. | Press |
| . |
|
|
| ||||||||
| Zero | ||||||||||||||
| Cal | ||||||||||||||
| 2. | Press |
|
|
|
|
|
|
| . During zeroing, which takes approximately 10 | |||||
|
|
|
|
|
|
| |||||||||
|
|
| Zero A |
| |||||||||||
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
| seconds, the wait symbol is displayed. | |||||||||||||
| 3. | Press |
| . The setting of the reference calibration factor is | |||||||||||
| Cal | ||||||||||||||
|
| obtained from the sensor calibration table and displayed under the | |||||||||||||
|
|
|
|
|
| softkey. | |||||||||
|
|
| A Ref CF | ||||||||||||
| 4. | Connect the power sensor to the POWER REF output. | |||||||||||||
| 5. | Press |
|
| to calibrate the power meter. During calibration the | ||||||||||
|
| Cal A | |||||||||||||
|
| wait symbol is displayed. (The power meter automatically turns | |||||||||||||
|
| on the POWER REF output.) |
HP |