Power Meter Operation
Calibrating the Power Meter
Calibration Procedure Using HP
The following procedure describes how you calibrate the power meter with an HP
Note | The following procedure details the calibration for channel A. To | |||||||
| calibrate channel B use the same procedure using the equivalent | |||||||
| channel B softkeys. | |||||||
|
|
|
|
|
|
|
|
|
| 1. | Press |
| . |
|
| ||
| Zero | |||||||
| Cal | |||||||
| 2. | Connect the power sensor to the POWER REF output. | ||||||
| 3. | Press |
|
| , |
| to calibrate the power meter. During | |
| Cal |
| Cal A | |||||
|
| calibration the wait symbol is displayed. (The power meter | ||||||
|
| automatically turns on the POWER REF output.) |
Example
To calibrate both channels of the power meter with HP
• Press Zero
Cal
,Cal .
•Connect the channel A power sensor to the POWER REF output.
•Press Cal A then wait till the calibration is completed.
•Remove the channel A power sensor from the POWER REF output.
•Connect the channel B power sensor to the POWER REF output.
•Press Cal B then wait till the calibration is completed.
Calibration Procedure using HP 8480 Series Power Sensors
The following procedure describes how you calibrate the power meter with the HP 8480 series power sensors.
There are a variety of different methods to connect the power sensors to the power meter depending on the model of power sensor you are using. Refer to Table
HP |