Power Meter Operation

Calibrating the Power Meter

Calibration Procedure Using HP E-Series Power Sensors

The following procedure describes how you calibrate the power meter with an HP E-series power sensor. Since the power meter automatically downloads the HP E-series power sensor’s calibration table there is no requirement to enter the reference calibration factor. The power meter identifies that an HP E-series power sensor is connected and will not allow you to select certain softkeys. The text on these softkeys appears grayed out.

Note

The following procedure details the calibration for channel A. To

 

calibrate channel B use the same procedure using the equivalent

 

channel B softkeys.

 

 

 

 

 

 

 

 

 

 

1.

Press

 

.

 

 

 

Zero

 

Cal

 

2.

Connect the power sensor to the POWER REF output.

 

3.

Press

 

 

,

 

to calibrate the power meter. During

 

Cal

 

Cal A

 

 

calibration the wait symbol is displayed. (The power meter

 

 

automatically turns on the POWER REF output.)

Example

To calibrate both channels of the power meter with HP E-series power sensors connected to both channels.

• Press Zero

Cal

,Cal .

Connect the channel A power sensor to the POWER REF output.

Press Cal A then wait till the calibration is completed.

Remove the channel A power sensor from the POWER REF output.

Connect the channel B power sensor to the POWER REF output.

Press Cal B then wait till the calibration is completed.

Calibration Procedure using HP 8480 Series Power Sensors

The following procedure describes how you calibrate the power meter with the HP 8480 series power sensors.

There are a variety of different methods to connect the power sensors to the power meter depending on the model of power sensor you are using. Refer to Table 2-1 on page 2-7 for details on connecting different power sensor models.

HP EPM-442A User’s Guide

2-5