Error Messages

Error Messages

-113

Undefined header

 

A command was received that is not valid for this power meter. You may

 

have misspelled the command or it may not be valid command. If you

 

are using the short form of the command, remember that it may contain

 

up to four letters.

 

For example, TRIGG:SOUR IMM.

-121

Invalid character in number

 

An invalid character was found in the number specified for a parameter

 

value.

 

For example, SENS:AVER:COUN 128#H.

-123

Exponent too large

 

A numeric parameter was found whose exponent was larger than

 

32,000.

 

For example, SENS:COUN 1E34000.

-124

Too many digits

 

A numeric parameter was found whose mantissa contained more than

 

255 digits, excluding leading zeros.

-128

Numeric data not allowed

 

A numeric value was received within a command which does not accept

 

a numeric value.

 

For example, MEM:CLE 24.

-131

Invalid suffix

 

A suffix was incorrectly specified for a numeric parameter. You may

 

have misspelled the suffix.

 

For example, SENS:FREQ 200KZ.

-134

Suffix too long

 

A suffix used contained more than 12 characters.

 

For example, SENS:FREQ 2MHZZZZZZZZZZZ.

-138

Suffix not allowed

 

A suffix was received following a numeric parameter which does not

 

accept a suffix.

 

For example, INIT:CONT 0Hz.

HP EPM-442A User’s Guide

4-5