Menu Reference
Front Panel Menu Reference "System/Inputs" Menu
■Done
Press this softkey to return to the “System/Inputs” menu.
•Recorder Output
Press this softkey to access a menu which allows you to modify the setup of the recorder output. Refer to “Recorder Output”, on page
■Output Off On
Press this softkey to switch the rear panel Recorder Output on or off. The Recorder Output produces a dc voltage that corresponds to the selected input channel power level in Watts.
■Max Power
Press this softkey to enter the input power level that you want to represent the 1 Vdc maximum output voltage of the Recorder Output.
■Min Power
Press this softkey to enter the input power level that you want to represent the 0 Vdc minimum output voltage of the Recorder Output.
■Output A B
Press this softkey to select which channel input signal appears on the Recorder Output.
•Service
Press this softkey to access a menu which allows you to test and service your power meter.
■
◆
◆
Self Test
Press this softkey to access the power meter’s self test menu.
Instrument Self Test
Press this softkey to run a series of tests on the power meter. Refer to “Instrument Self Test”, on page
Confidence Check
Press this softkey to verify that the power meter makes an accurate measurement of it’s 1 mW POWER REF output.
◆Individual
Press this softkey to access a menu which allows you to select individual self tests to be run.