Frequency Cal Fac

Power Meter Operation

Making Measurements with the HP 8480 Series Power Sensors

6.The current setting of the reference calibration factor is displayed under the A Ref CF softkey. To change this setting press

A Ref CF . The power meter displays the reference calibration factor in a pop up window. Modify this reference calibration factor (see below) as desired.

Use or to modify the digit on which the cursor is currently positioned.

Use or to move to other digits.

7.To confirm your choice press % .

8.Press Cal, Cal A to calibrate the power meter. During calibration the wait symbol is displayed. (The power meter automatically turns on the POWER REF output.)

9. Press . The current setting of the calibration factor is displayed under the A Cal Fac softkey. To change this setting press A Cal Fac. The power meter displays the calibration factor in a pop up window. Modify this calibration factor (see below) as desired.

Use or to modify the digit on which the cursor is currently positioned.

Use or to move to other digits.

10.To confirm your choice press % .

11.Connect the power sensor to the signal to be measured.

12.You must set the display to the type of measurement you require. This can either be a direct channel measurement, a ratio measurement of both channels, or a difference measurement

between both channels. Use the following key presses:

Press SetupMeas , Input Select. Select the appropriate softkey for the measurement you require, either A, B, A/B , B/A , or More , then A-B , or B-A .

13.The measurement result is now displayed.

Example

To make a measurement on channel A with a power sensor which has a reference calibration factor of 99.8% and a calibration factor of 97.8% at the measurement frequency.

Disconnect the power sensor from any source.

Press ZerCalo .

Press Zero A.

Press Cal.

HP EPM-442A User’s Guide