Power Meter Operation

Calibrating the Power Meter

 

Calibrating the Power Meter

 

This section describes how to calibrate the power meter. You should

 

always zero the power meter prior to calibrating it.

 

Calibration sets the gain of the power meter using a 50 MHz 1 mW

 

calibrator as a traceable power reference. The power meter’s POWER REF

 

output or a suitable external reference is used as the signal source for

 

calibration. An essential part of calibrating is setting the correct reference

 

calibration factor for the power sensor you are using. The HP 8480 series

 

power sensors require you to set the reference calibration factor. The

 

HP E-series power sensors set the reference calibration factor

 

automatically. During calibration the wait symbol is displayed. Offset,

 

relative and duty cycle settings are ignored during calibration.

 

 

Note

During calibration the power meter automatically switches the

 

power reference calibrator on (if it is not already on), then after

 

calibration it switches it to the state it was in prior to the

 

calibration.

 

 