Menu Reference
Front Panel Menu Reference "System/Inputs" Menu
■
■
■
■
■
entered using Filter . Refer to “Setting Averaging”, on page
Duty Cycle Off On
Press this softkey to toggle the duty cycle value on or off. The default is “Off”. The duty cycle value is set using
Duty Cycle .
Duty Cycle
Press this softkey to set the duty cycle for the pulse power measurement feature of the power meter. A value in the range of 0.001% to 100% can be entered. The default value is 1.000%.
Use the , , and hardkeys to change the value. To confirm your choice press % . Duty Cycle Off On is automatically set to “On” when a value is entered using Duty Cycle . Refer to “Measuring Pulsed Signals”, on page
Offset Off On
Press this softkey to toggle the channel offset value on or off. The default is “Off”. The offset value is set using Offset.
Offset
Press this softkey to enter a channel offset in the range of
, and hardkeys to change the value. To confirm your choice press dB. This offset can be used to compensate for a loss or gain. The offset is applied to the measured power before the result is displayed. Offset Off On is automatically set to “On” when a value is entered using Offset . Refer to “Setting Channel Offsets”, on page
Range (HP
Press this softkey to set the range in which you require the power meter to make the power measurement or set the power meter to autoranging. Choose either “UPPER”, “LOWER” or
“AUTO” from the selection available using the and hardkeys. The default is “AUTO”. Refer to “Setting the Range”, on page
■Limits
Press this softkey to access a menu which allows you to enter the upper and lower measurement limits. Refer to “Setting Measurement Limits”, on page