Power Meter Operation

Saving and Recalling Power Meter Configurations

Saving and Recalling Power Meter Configurations

To reduce repeated setup sequences, up to ten power meter configurations can be stored in the power meter’s non-volatile memory. The HP-IB address, the data stored in the sensor calibration tables and the zeroing and calibration data are not stored by a preset. The calibration table selected is not affected.

Power meter configurations are saved and recalled with the hardkey.

To save the current measurement setup:

1. Press SaveRecall .

Save

Recall

2.Use the and hardkeys to scroll through the displayed files. When the required file is highlighted press Save .

3.Press Confirm .

If you require to modify the name of a file:

1.Press SaveRecall .

2.Use the and hardkeys to scroll through the displayed files. When the required file is highlighted press Edit Name . The power meter displays the file name in a pop up window. Modify this name (see below) until the desired name is displayed.

Use or to modify the character on which the cursor is currently positioned.

Use or to move to other characters.

Use Insert Char and Delete Char as required.

3.To confirm your choice press Enter .

To recall a measurement setup:

1.Press SaveRecall .

2.Use the and hardkeys to scroll through the displayed files. When the required file is highlighted press Recall .

3.Press Confirm .

HP EPM-442A User’s Guide